Concurrent Validity and Within-Session Reliability of a Wireless Surface Electromyography Device (MR EMG) Compared to a Criterion Measure During the Leg Extension Exercise.
Jamie J Ghigiarelli, Adam M Gonzalez, Grace A Valdez, Thomas L Johannessen, Admar L Calovini, James M Hackney
Department of Allied Health and Kinesiology, Hofstra University, Hempstead, NY 11549, USA.
Key points
From the paper's conclusion- Purpose: The aim of this study was to assess within-session reliability and concurrent validity of a wireless surface electromyography device (MR EMG) relative to a wireless criterion system (Delsys Trigno) during dynamic leg extension at varying loading intensities.
Abstract
Purpose: The aim of this study was to assess within-session reliability and concurrent validity of a wireless surface electromyography device (MR EMG) relative to a wireless criterion system (Delsys Trigno) during dynamic leg extension at varying loading intensities. Methods: Thirty-one resistance-trained adults performed four repetitions at 30%, 60%, and 90% of their predicted one-repetition maximum (1RM) on the plate-loaded leg extension machine, with EMG amplitude recorded simultaneously by both systems for the vastus lateralis (VL) and vastus medialis obliquus (VMO). Root mean square (RMS) amplitudes were normalized to maximal voluntary isometric contraction and compared across devices using intraclass correlation coefficients (ICC), coefficients of variation (CV%), standard error of measurement, Pearson correlations, and Bland-Altman analyses. Results: Both devices demonstrated excellent within-session reliability (ICC = 0.92-0.97; CV = 7.8-11.3%). Validity differed by muscle. Across all loads, VMO agreement was good (ICC = 0.82-0.85; r = 0.83-0.86), while VL agreement was moderate (ICC = 0.53-0.57; r = 0.54-0.64). Bland-Altman analyses revealed systematic positive bias across all conditions, with proportional bias present for the VL but absent for the VMO. Conclusions: MR EMG demonstrated excellent within-session reliability but muscle-dependent concurrent validity, with good VMO agreement and moderate VL agreement with the criterion system.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- How Stable Are Temporal EMG Parameters in Rowing? A Seven-Day Test-Retest Reliability Study Using Wearable sEMG.Sensors (Basel, Switzerland) · 2026
- Surface-EMG Amplitude Differences Before and After Scheduled University Team Training Following a Standardized Warm-Up in Male Collegiate Volleyball Players.Journal of clinical medicine · 2026
- Whole-repetition quadriceps sEMG during flywheel squat and leg press: a randomized crossover study.Frontiers in physiology · 2026
- Inter-day ranking of biceps brachii sEMG amplitude during four arm curl exercises.Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology · 2026
- Relationship between rectus femoris stiffness, muscle activity, and force production during graded isometric contractions measured by handheld myotonometry.Journal of sports sciences · 2026
- Comparing Repetition-Velocity Output from a Motor-Integrated Magnetic Sensing System with That of a Linear Position Transducer Across Resistance Exercises and Loads: A Human Trial Experiment.Sensors (Basel, Switzerland) · 2026