Characterization of Silicon-Membrane TES Microcalorimeters for Large-Format X-ray Spectrometers with Integrated Microwave SQUID Readout.
Avirup Roy, Robinjeet Singh, Joel C Weber, W B Doriese, Johnathon Gard, Mark W Keller, John A B Mates, Kelsey M Morgan,
Department of Physics, University of Colorado Boulder, Boulder, CO 80309 USA. · National Institute of Standards and Technology, Boulder, CO 80305 USA.
Key points
From the paper's conclusion- This work is motivated by our goal to develop a 10,000-pixel TES spectrometer to overcome critical measurement limitations in catalysis research.
Abstract
We present the electro-thermal characterization of transition-edge sensor (TES) detectors suspended on Si membranes fabricated using a silicon-on-insulator (SOI) wafer. The use of an all-silicon fabrication platform, in contrast to the more commonly used silicon nitride membranes, is compatible with monolithic fabrication of integrated TES and SQUID circuits. The all-silicon architecture additionally allows efficient use of focal plane area; the readout circuitry may be positioned out of the focal plane by bending a thinned portion of the chip. Compatibility with integrated fabrication and efficient use of focal plane area provide a path to an efficient soft X-ray spectrometer. This work is motivated by our goal to develop a 10,000-pixel TES spectrometer to overcome critical measurement limitations in catalysis research. The characterization of fragile, carbon-based intermediates via techniques like Resonant Inelastic X-ray Scattering (RIXS) is often precluded by the slow, high-flux nature of existing technologies. The new instrument will allow for fast RIXS measurements to be made without causing sample damage. We verify the detector models and measure the energy resolution using a pulsed optical laser, demonstrating the viability of this approach for the final instrument to be deployed at the National Synchrotron Light Source II (NSLS-II).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 2D-RIXS: resonant inelastic X-ray scattering microscopy with high energy and spatial resolutionsJournal of Synchrotron Radiation · 2026
- Crystalline Tellurium Nanowires Enabled by Rapid Hydrolysis of Lithium-Intercalated Layered Metal Tellurides.Small (Weinheim an der Bergstrasse, Germany) · 2026
- Synergistic Time‐Delay Integration X‐Ray Imaging with Pixelated Cs 3 Cu 2 I 5 Scintillators for High‐Fidelity In‐Line InspectionAdvanced Optical Materials · 2026
- Controlled Oxidation of Tellurium Nanoparticles for Telecom DetectionarXiv · 2026
- Interfacial Modulation via Ultrathin ALD‐MgO for High‐Performance SnO 2 /Te 0.75 Se 0.25 Short‐Wave Infrared PhotodetectorsAdvanced Optical Materials · 2026
- Optical Calibration of a Frequency-Domain-Multiplexed X-Ray TES Microcalorimeter Array via Photon-Number-Resolved Statistics.Sensors (Basel, Switzerland) · 2026