Clinical and Biomechanical Evaluation of an Adaptive Polyarticulated Myoelectric Hand Prosthesis: A Case Report
Mattia Giuseppe Viva, Alessio Fricano, Marco Paoloni, Andrea Bernetti, Giuseppe Giannicola, Mario Vetrano, Francesco Agostini, Giovanni Galeoto,
Key points
From the paper's conclusion- Conclusions: This case highlights a task-dependent biomechanical profile of an adaptive polyarticulated myoelectric hand under first-use conditions.
Abstract
Introduction: Adaptive polyarticulated myoelectric hands aim to improve grasp versatility and intuitiveness of use; however, objective clinical and biomechanical evidence supporting their functional impact remains limited. Case presentation: A 75-year-old right-hand-dominant male with a long-standing transradial traumatic amputation and over 30 years of experience with a myoelectric tridigital prosthesis underwent experimental evaluation of an adaptive polyarticulated myoelectric hand (Adam’s Hand®). Methods: Assessment included patient-reported outcome measures (Orthotics and Prosthetics User Survey—Upper Extremity Functional Status (OPUS-UEFS), Trinity Amputation and Prosthesis Experience Scales–Revised (TAPES-R), Disabilities of the Arm, Shoulder and Hand questionnaire (DASH), Short Form Health Survey 36 (SF-36)), standardized performance-based functional tests (Minnesota Manual Dexterity Test (MMDT), Box and Block Test (BBT), Southampton Hand Assessment Procedure (SHAP)), and laboratory-based biomechanical evaluation using optoelectronic motion capture and surface electromyography during Pointing, Hand-to-Mouth, and Grasping tasks. The adaptive prosthesis was tested after rapid artificial intelligence-based calibration, without structured training. Results: Patient-reported outcomes reflected high satisfaction and adaptation to the habitual prosthesis. Performance-based tests favored the habitual device in speed and dexterity, whereas the adaptive hand enabled additional grasp patterns. Biomechanically, the adaptive hand was smoother and faster during reaching, but less stable during grasping. Compensatory trunk and shoulder movements occurred with both devices, with a trend toward less compensation during reaching with the adaptive hand. Conclusions: This case highlights a task-dependent biomechanical profile of an adaptive polyarticulated myoelectric hand under first-use conditions. Rapid calibration may support immediate use for reaching, while optimal grasp control appears to require structured training.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Phase-Based Gesture Motor Learning Protocol for Training Prosthetic Control in Individuals with Transradial AmputationMethods and Protocols · 2026
- Estimation of Endurance During Elbow Flexion With Myoelectric Orthotic AssistanceArchives of Physical Medicine and Rehabilitation · 2026
- An Adaptive Multi-Representation Fusion Framework for EMG Pattern Recognition in Transradial Amputees.Bioengineering (Basel, Switzerland) · 2026
- Multimodal Feedback-Enhanced Soft Hand Rehabilitation Robot: System Development and Preliminary Clinical Evaluation.Bioengineering (Basel, Switzerland) · 2026
- Onset-Triggered Vote-Locking (OTVL) for Stable Myoelectric Control of Multi-DoF Prosthetic Hands.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- Mitigating Muscle Fatigue in Upper-Limb Prosthesis Users Through Exoskeletal Weight CompensationIEEE Transactions on Neural Systems and Rehabilitation Engineering · 2025