Influence of Update Rate on the Susceptibility of Standoff Open-Path TDLAS Measurements to Drone-Induced Mechanical Noise
Giacomo Zanetti, Peter John Rodrigo, Christian Pedersen
Key points
From the paper's conclusion- We observed that increasing the measurement frequency reduced the integration time required to freeze the scene, but the limited speed and resolution of the electronics used meant an optimal working point of 15.43 kHz for our configuration.
Abstract
Tunable diode laser absorption spectroscopy (TDLAS) is a well-established spectroscopic method used for remote gas concentration measurements in real-world environments. However, mechanical vibrations affecting optical components exposed to environmental conditions can limit the precision of these measurements. In this work, we experimentally demonstrate that precompensating the current modulation to enable rapid switching between two or three wavelengths reduces the integration time required to achieve a “frozen scene,” i.e., a state effectively free of mechanical vibration effects. This improvement was demonstrated by analyzing the overlapping Allan deviation of CO2 measurements under atmospheric conditions (approximately 430 ppm) as a function of the instrument’s update rate, which ranged from 156 Hz to 30.86 kHz, achieving the fastest real-time remote TDLAS concentration measurements reported to date. The experimental setup included a drone-mounted retroreflector, and the Allan deviation was characterized both with the drone turned off and on while clamped to a stage. We observed that increasing the measurement frequency reduced the integration time required to freeze the scene, but the limited speed and resolution of the electronics used meant an optimal working point of 15.43 kHz for our configuration. Lastly, this characterization served to select the update rate to utilize during an outdoor measurement which confirmed the usefulness of the high update rates in the suppression of the effect of drone-induced mechanical noise on concentration measurements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Enhancing High‐Frequency Fidelity of Open‐Path TDLAS Sensors by Dynamic Background Noise SuppressionMicrowave and Optical Technology Letters · 2026
- Full-range CO2 detection using dual-path WMS-DAS and multi-regime adaptive fusion.Optics letters · 2026
- Continuous-streaming high-speed two-photon dual-comb LiDAR with free-running lasers.Optics express · 2026
- Simplified Calibration and Laser Nonlinearity Error Analysis for Peak-Locking TDLAS SystemsPhotonics · 2026
- Baseline-Free Wide-Range Methane Concentration Measurement Based on Logarithmic Second Derivative Spectroscopy.Analytical chemistry · 2026
- Dynamic calibration method for the system constant of high-spectral-resolution lidar with airborne experimental verification.Optics express · 2026