A Simple and Green LC–MS/MS Method for the Simultaneous Quantification of Clarithromycin, Its Two Metabolites, and Loxoprofen in Human Plasma: Application to a Pharmacokinetic Study
Sangyoung Lee, Sabin Shin, Da Hyun Kim, Jee Sun Min, Chae Bin Lee, Duk Yeon Kim, Ui Min Jerng, Soo Kyung Bae
Key points
From the paper's conclusion- To support a clinical pharmacokinetic study in which both drugs were co-administered, this study aimed to develop and validate a simple and environmentally sustainable LC–MS/MS method for the simultaneous quantification of CLA, its two metabolites 14-hydroxy clarithromycin (14-OH) and N-desmethyl clarithromycin (NDM), and LOX in human plasma.
Abstract
Background/Objectives: Clarithromycin (CLA), a macrolide antibiotic, and loxoprofen (LOX), a nonsteroidal anti-inflammatory drug, may be concomitantly prescribed for respiratory tract infections. To support a clinical pharmacokinetic study in which both drugs were co-administered, this study aimed to develop and validate a simple and environmentally sustainable LC–MS/MS method for the simultaneous quantification of CLA, its two metabolites 14-hydroxy clarithromycin (14-OH) and N-desmethyl clarithromycin (NDM), and LOX in human plasma. Methods: A 50 μL aliquot of human EDTA plasma was processed via acetonitrile protein precipitation, using chlorpropamide as the internal standard (IS). Chromatographic separation was achieved on a C18 column under gradient elution with acetonitrile and ultrapure water, both containing 0.1% formic acid, at a flow rate of 0.3 mL/min. Detection was performed using positive-mode electrospray ionization with multiple-reaction monitoring at m/z 748.8 > 158.2, 764.7 > 158.0, 734.5 > 144.4, 247.1 > 201.1, and 277.1 > 175.3 for CLA, 14-OH, NDM, LOX, and the IS, respectively. Results: Calibration curves were linear (r ≥ 0.9915) over the ranges of (ng/mL) 10–4500 for CLA, 10–2000 for 14-OH, 2–1500 for NDM, and 50–20,000 for LOX. The method was fully validated in EDTA plasma and partially validated in lithium–heparin and sodium–heparin plasma. All validation parameters met the acceptance criteria defined in the ICH M10 guideline. Method greenness was assessed using the AGREE (0.64) and Complex MoGAPI (78) metrics. Conclusions: The validated method was successfully applied to plasma samples from healthy volunteers following oral co-administration of 250 mg CLA and 60 mg LOX, enabling the pharmacokinetic characterization of all four analytes from a single 50 μL sample; its applicability to other populations remains to be established.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Organic solvent-free micellar liquid chromatographic determination of amoxicillin, chlorzoxazone, and ibuprofen in pharmaceutical and biological matrices.Scientific reports · 2026
- A robust liquid chromatography–tandem mass spectrometry method for accurate quantification of omeprazole in human plasma: Integration of International Council for Harmonisation guidelines for pharmacokinetic applicationActa Chromatographica · 2026
- Sustainable UHPLC-MS/MS assay for ultra-sensitive simultaneous quantification of amoxicillin and flucloxacillin with application to a bioequivalence study.Analytical and bioanalytical chemistry · 2026
- Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.Journal of separation science · 2026
- Quantification of clofazimine, linezolid, levofloxacin, and moxifloxacin in breast milk by online solid phase extraction coupled to LC-MS/MS.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Development and validation of a LC-MS/MS method for the quantification of the E-type prostanoid receptor 4 antagonist HTL0039732 in human plasma, for a Phase I/IIa clinical trial.Bioanalysis · 2026