Environmentally Friendly LC-MS/MS Method for the Determination of Etrasimod in Human Plasma via Advanced Phospholipid Matrix Removal.
Jelena Kovačić Šantek, Daniela Amidžić Klarić, Nikša Turk, Željko Krznarić, Ana Mornar
Department of Pharmaceutical Analysis, Faculty of Pharmacy and Biochemistry, University of Zagreb, A. Kovačića 1, 10 000 Zagreb, Croatia. · Department of Gastroenterology, University Hospital Centre, Kišpatićeva 12, 10 000 Zagreb, Croatia.
Key points
From the paper's conclusion- Conclusions: The proposed method provides a reliable and green tool suitable for analysis of etrasimod in human plasma.
Abstract
Background/Objectives: Etrasimod is a novel, oral, selective sphingosine-1-phosphate (S1P) receptor modulator approved for the treatment of ulcerative colitis. Due to its high lipophilicity and strong plasma protein binding, quantification of etrasimod in human plasma is significantly challenged by co-extracted proteins and phospholipids that induce column degradation and severe mass spectrometry matrix effects. This study presents the first eco-friendly LC-MS/MS assay for etrasimod in human plasma utilizing advanced phospholipid removal technology, combining protein precipitation and targeted phospholipid depletion into a single step. It achieved matrix purification while reducing sample processing time and solvent volumes. Methods: Chromatographic separation was achieved on a C18 column using a gradient elution of 0.1% formic acid in water and acetonitrile. Detection was performed on a triple quadrupole mass spectrometer in positive electrospray ionization (ESI) mode using multiple reaction monitoring (MRM) transitions at m/z 458 → 159 for etrasimod and m/z 405 → 146 for ozanimod (internal standard). Results: The method was fully validated according to ICH M10 guidelines, demonstrating excellent linearity, accuracy, precision, and robustness. Comprehensive sustainability assessment using AGREE (0.55), AGREEprep (0.58), ComplexMoGAPI (78), and BAGI (52.5) confirmed an eco-friendly profile and high operational practicality. Conclusions: The proposed method provides a reliable and green tool suitable for analysis of etrasimod in human plasma.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A simplified EMR-Lipid cleanup strategy coupled with UPLC-MS/MS for determination of legacy and emerging PFAS in eggs: method validation and greenness assessment.Analytical methods : advancing methods and applications · 2026
- A multi-modal mass spectrometric approach for lipid analysis enabled by online fractionation of complex mixtures in a solid-phase extraction cartridge.The Analyst · 2026
- Development and validation of a LC-MS/MS method for the quantification of the E-type prostanoid receptor 4 antagonist HTL0039732 in human plasma, for a Phase I/IIa clinical trial.Bioanalysis · 2026
- SIL-IS LC-MS/MS method for rapid quantification of phloroglucinol in human plasma: development, validation and application to a pharmacokinetic study.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- TiO2-modified paper spray mass spectrometry for selective phospholipid removal in milk and plasma.Talanta · 2026
- A TiO2@MWCNTs-based dispersive solid-phase extraction method for mitigating matrix effects in plasma LC-MS/MS analysis.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026