Left Atrial Appendage Occlusion with the Watchman FLX Device: A Single-Centre Retrospective Analysis of Procedural and One-Year Clinical Outcomes.
Wasim Allakkis, Mostafa Salem, Mohamed Salem, Mohamed Elabd, Ali Balakishiyev, Moussab Aboud, Ammar Fayoud, Mohammed Saad,
Department of Cardiology, Helios Frankenwaldklinik-Hospital, 96317 Kronach, Germany. · Department of Cardiology, University Hospital Schleswig-Holstein, 24105 Kiel, Germany.
Key points
From the paper's conclusion- This single-centre retrospective study describes device implantation, peri-procedural complications and clinical outcomes during the first year after Watchman FLX implantation.
Abstract
Background: Left atrial appendage occlusion (LAAO) is an alternative approach to stroke prevention in selected patients with non-valvular atrial fibrillation. This single-centre retrospective study describes device implantation, peri-procedural complications and clinical outcomes during the first year after Watchman FLX implantation. Methods: We analysed 100 consecutive patients who underwent LAAO with the Watchman FLX device at Helios Frankenwaldklinik-Hospital, Kronach, Germany, between 2021 and 2024. The primary endpoint was device implantation success, defined as correct device deployment and implantation. The secondary safety endpoint was a composite of major peri-procedural complications within 72 h. Additional clinical outcomes were strokes and mortality at 30 days and one year. Results: The mean age was 70.7 ± 6.5 years, and 47% of patients were female. The mean CHA2DS2-VASc score was 4.9 ± 1.4, and the mean HAS-BLED score was 4.1 ± 1.0. Device implantation was successful in 100% (95% CI 96.4-100%). The reported peri-procedural complication rate was 7% (95% CI 2.9-13.9%), including major bleeding in 6% (95% CI 2.2-12.6%) and pericardial tamponade in 1% (95% CI 0.0-5.4%). Thirty-day mortality was 2% (95% CI 0.2-7.0%), with no reported strokes (0%; 95% CI 0.0-3.6%). One-year mortality was 11% (95% CI 5.6-18.8%), with one reported stroke (1%; 95% CI 0.0-5.4%). Conclusions: Watchman FLX implantation was achieved in all patients in this cohort. Bleeding was the most frequently reported peri-procedural complication. The low observed stroke proportion should be interpreted alongside the 11% one-year mortality and incomplete imaging surveillance. This uncontrolled descriptive study cannot establish comparative efficacy or comprehensive device safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world complication profiles of left atrial appendage occlusion devices: a MAUDE database analysis of 8,406 adverse eventsEuropean Heart Journal · 2026
- Incidence, Predictors, Management, and Clinical Impact of Major Bleeding Following Left Atrial Appendage OcclusionJAMA Cardiology · 2026
- Persistent peri-device leak following left atrial appendage occlusion treated with hybrid coil- and plug-based closure: a case report.European heart journal. Case reports · 2026
- Three-year clinical outcomes of left atrial appendage occlusion with or without ablation: insight from the RECORD studyEP Europace · 2026
- Watching 'AF'ter the Heart: the WATCHMANTM device for atrial fibrillation.Future cardiology · 2026
- Left atrial appendage closure: Current status, unresolved issues, and future perspectivesCardiovascular Intervention and Therapeutics · 2025