Materials (Basel, Switzerland)

Scintillation Fiber-Optic Detectors for Dosimetry of 60 MeV Proton Beam.

Sandra Witkiewicz-Lukaszek, Bogna Sobiech, Paweł Bilski, Anna Mrozik, Michał Sądel, Jan Swakoń, Damian Wróbel, Janusz Winiecki,

Faculty of Physics, Kazimierz Wielki University in Bydgoszcz, Powstańców Wielkopolskich Street 2, 85-090 Bydgoszcz, Poland. · The Franciszek Łukaszyk Oncology Center, Medical Physics Department, Dr. Izabeli Romanowskiej Street 2, 85-796 Bydgoszcz, Poland.

Key points

From the paper's conclusion
  • Overall, these results demonstrate that scintillation FODs constitute a robust platform for proton beam diagnostics and dosimetry.

Abstract

Scintillation fiber-optic detectors (FODs) offer a compact, electrically passive, and versatile solution for real-time proton beam monitoring in clinical radiation therapy. In this work, we systematically investigate and compare FODs based on GAGG:Ce, Al2O3:C, and Al2O3:C,Mg crystal scintillators under identical proton irradiation conditions. The detectors were evaluated using a 60 MeV clinical proton beam over a dose range of 0.5-15 Gy, with radioluminescence spectra acquired at integration times as short as 70 ms. The GAGG:Ce-based FOD exhibited the highest signal intensity, a fast temporal response, and excellent stability under repeated irradiation cycles, enabling reliable real-time dose monitoring. In contrast, the FODs based on Al2O3:C and Al2O3:C,Mg demonstrated high sensitivity, reproducible dose-response characteristics, and a low effective atomic number (Zeff ≈ 11), providing near tissue-equivalent behavior. Mg co-doping modified the luminescence characteristics of Al2O3:C by enhancing the contribution of visible emission centers. However, under the present proton irradiation conditions, the overall RL signal remained lower than that of Al2O3:C while preserving good signal stability. All detector configurations maintained consistent performance across multiple irradiation cycles, with GAGG:Ce showing the lowest signal variation. Overall, these results demonstrate that scintillation FODs constitute a robust platform for proton beam diagnostics and dosimetry. GAGG:Ce-based detectors are shown to have promising potential for real-time proton beam monitoring, whereas sapphire-based detectors provide advantages for dosimetric applications where tissue equivalence is essential.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers