Comparative Assessment of Pre-Dialysis Fluid Status Using Clinical, Lung, and IVC Ultrasound Methods with Bioimpedance Spectroscopy as Reference Standard in Hemodialysis Patients
Ajith M. Nayak, Attur Ravindra Prabhu, Indu Ramachandra Rao, Mohan V. Bhojaraja, Shwetha Prabhu, Bharathi Naik, Dharshan Rangaswamy, Srinivas Vinayak Shenoy,
Key points
From the paper's conclusion- Conclusions: LUS had the highest observed AUC; nephrologist’s clinical classification missed more BCM-defined overhydrated patients than LUS or IVC ultrasound.
Abstract
Background: Accurate assessment of overhydration in maintenance hemodialysis (MHD) patients remains challenging. The clinical method is widely used but may be unreliable. This study compared the nephrologist’s clinical method, lung ultrasound (LUS), and inferior vena cava (IVC) ultrasound against bioimpedance spectroscopy using a body composition monitor (BCM). Methods: In this single-center cross-sectional study, 98 hypertensive patients receiving twice-weekly MHD were evaluated within 30 min before the second dialysis session of the week. LUS B-lines, IVC diameter index at expiration (IVCDimax), and IVC collapsibility index (IVCCi) were measured. BCM-defined overhydration (>2.5 L) was the reference standard. Results: BCM classified 88/98 (89.8%) patients as overhydrated; among these, LUS, IVC ultrasound, and nephrologist’s clinical classification identified 86, 82, and 65 patients, respectively. LUS B-lines showed the strongest correlation with BCM-defined overhydration (ρ = 0.913) and the highest observed AUC (0.985; exploratory cutoff ≥12 B-lines: sensitivity 97.7%, specificity 90.0%). IVCCi showed a strong inverse correlation (r = −0.830; AUC = 0.940), while IVCDimax showed a moderate correlation (ρ = 0.641; AUC = 0.853). Clinically estimated overhydration showed a strong correlation (r = 0.797) but lower discriminative ability (AUC = 0.679) and a mean bias of −0.887 L. Conclusions: LUS had the highest observed AUC; nephrologist’s clinical classification missed more BCM-defined overhydrated patients than LUS or IVC ultrasound.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Physiological discordance between clinical examination, bioimpedance spectroscopy, and lung ultrasound in maintenance hemodialysis patients without heart failure.Frontiers in nephrology · 2026
- Combined ultrasonographic and clinical assessment of bioimpedance-defined volume status in dialysis patients: a diagnostic accuracy study.Journal of ultrasound · 2026
- Discordance between extracellular volume and hemodynamic congestion in hemodialysis: A pilot multiparametric study.Nefrologia · 2026
- Ultrasound-guided haemodynamic profiling in haemodialysis patients: a single-centre prospective observational study protocol.Frontiers in nephrology · 2026
- Multimodal Diagnostic Ultrasound for Congestion, Perfusion, and Ultrafiltration Tolerance in Maintenance Hemodialysis: A Narrative Review.Diagnostics (Basel, Switzerland) · 2026
- Concordance of a Laboratory-Data Integrated Fluid Targeting Algorithm with Bioimpedance-Derived Volume Measures in Maintenance Hemodialysis.Kidney360 · 2026