Quantification of Fosfomycin in Plasma by LC-MS/MS: Development, Validation and Application to Pharmacological Monitoring.
Alice Palermiti, Camilla Martino, Maria Donata Valenti, Miriam Antonucci, Sara Soloperto, Martina Billi, Gianluca Bianco, Riccardo Marras,
Laboratory of Clinical Pharmacology and Pharmacogenetics, Department of Medical Sciences, University of Turin, 10149 Turin, Italy. · ASL Città di Torino, Amedeo di Savoia Hospital, 10149 Turin, Italy.
Key points
From the paper's conclusion- Validation according to the International Council for Harmonisation (ICH) M10 guideline demonstrated linearity over 1.56–200 mg/L (lower limit of quantification of 1.56 mg/L; limit of detection 0.19 mg/L), intra- and inter-day precision and accuracy within acceptance criteria acceptable recovery and matrix effect in six independent plasma lots, and no relevant endogenous interference.
Abstract
Fosfomycin is an antibiotic of growing interest for the treatment of infections caused by multidrug-resistant bacteria. Its marked pharmacokinetic variability, especially in critically ill patients, makes therapeutic drug monitoring (TDM) useful to ensure adequate exposure and to limit the risk of treatment failure, resistance development, and toxicity. Given the high polarity of fosfomycin, a hydrophilic interaction liquid chromatography- tandem mass spectrometry (HILIC-MS/MS) method for its quantification in human plasma was developed and validated, using a Phenomenex Luna Polar Pesticides column, a Shimadzu LCMS-8050 mass spectrometer operated in negative electrospray ionisation mode and fosfomycin-d5 as internal standard. Sample preparation relied on a single-step protein precipitation of 40 µL of plasma, with a retention time of 2.8 min and a total run time of seven min. Validation according to the International Council for Harmonisation (ICH) M10 guideline demonstrated linearity over 1.56-200 mg/L (lower limit of quantification of 1.56 mg/L; limit of detection 0.19 mg/L), intra- and inter-day precision and accuracy within acceptance criteria acceptable recovery and matrix effect in six independent plasma lots, and no relevant endogenous interference. The method was applied to 44 samples from six patients receiving intravenous fosfomycin, and incurred sample reanalysis met ICH M10 acceptance criteria. This method provides a sensitive, selective, and robust tool for fosfomycin TDM pharmacokinetic studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Inhibition of Fosfomycin Resistance Protein FosA from Pseudomonas aeruginosa.ACS infectious diseases · 2026
- An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of piperacillin in human serum and plasma.Clinical chemistry and laboratory medicine · 2026
- High‐Sensitivity LC–MS/MS Method for Simultaneous Determination of Structurally Related Impurities 2‐Isoniazid, 3‐Isoniazid and Isonicotinamide in Isoniazid Active Pharmaceutical IngredientJournal of Mass Spectrometry · 2026
- Comprehensive evaluation of analytical methods for determination of ansamycins across diverse matrices.Analytical methods : advancing methods and applications · 2026
- SIL-IS LC-MS/MS method for rapid quantification of phloroglucinol in human plasma: development, validation and application to a pharmacokinetic study.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Development and validation of a novel LC-MS/MS method for simultaneous quantification of fidaxomicin and metabolite (OP-1118) from feces for gut pharmacobiome studies.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026