A Simplified LC-MS/MS Method for the Determination of Anticoagulant Rodenticides in Emergency Department Patients with Unexplained Coagulopathy.
Hsiao-Chia Liao, Shu-Mei Chi, Kuan-Ling Chen, Jyun-Han Lin, Su-Chin Chen, Yu-Cheng Li, Szu-Yin Hsieh, Shih-Rong Lin,
Department of Medical Toxicology, Taichung Veterans General Hospital, Taichung 407219, Taiwan. · SCIEX, Taipei 114063, Taiwan.
Key points
From the paper's conclusion- We developed and validated a simplified liquid chromatography–tandem mass spectrometry (LC-MS/MS) method for the simultaneous quantification of five ARs and warfarin in human plasma and evaluated its clinical applicability in a patient with unexplained coagulopathy.
Abstract
Anticoagulant rodenticide (AR) poisoning may cause prolonged and potentially life-threatening coagulopathy, while confirmation of exposure can be challenging when the exposure history is unclear. We developed and validated a simplified liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous quantification of five ARs and warfarin in human plasma and evaluated its clinical applicability in a patient with unexplained coagulopathy. The method used 50 μL of plasma with a simple protein-precipitation procedure followed by LC-MS/MS analysis. The validated quantitative range was 10-1000 ng/mL, with a lower limit of quantification of 10 ng/mL, while all analytes fulfilled the predefined qualitative identification criteria at 5 ng/mL. The method demonstrated acceptable accuracy, precision, selectivity, and matrix-effect performance. In the clinical case, bromadiolone was identified, and two plasma measurements showed a declining concentration profile, with an estimated apparent elimination half-life of 10.4 days. Based on this estimate, the plasma bromadiolone concentration was predicted to decline below 10 ng/mL at approximately day 42. This estimated concentration trajectory, together with the patient's clinical course and serial coagulation findings, supported the decision to discontinue oral vitamin K1 on day 42. Subsequent INR measurements remained normal on days 45 and 49. Although 10 ng/mL served as a practical reference in this case, it should not be used as the sole criterion for vitamin K1 discontinuation. This simplified LC-MS/MS method provides a practical approach for identifying and quantifying AR exposure in patients with otherwise unexplained coagulopathy and may support clinical assessment and management.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Forensic LC-MS/MS Monitoring of Anticoagulant Rodenticide Poisoning in Cats Over a Half-Decade.Veterinary medicine and science · 2026
- Tracking the Fall in Prothrombin Time in Dogs With Anticoagulant Rodenticide Toxicosis Following an Intravenous Dose of Mixed Micelle Phytomenadione.Journal of veterinary emergency and critical care (San Antonio, Tex. : 2001) · 2026
- Severe fetal hemorrhage and maternal coagulopathy following intentional bromadiolone ingestion in late pregnancy: a case report.International journal of surgery case reports · 2026
- Development and Validation of a QuEChERS-LC-MS/MS Method for Simultaneous Determination of Four Direct Oral Anticoagulants in Citrate Plasma and Quantitative Dried Plasma Spots.Journal of separation science · 2026
- A Validated Multi-Analyte Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Therapeutic Drug Monitoring and Targeted Toxicological Assessment of 54 Drugs and Metabolites in Human Serum.Journal of separation science · 2026
- Bridging of Finger-Prick Volumetric Absorptive Microsampling and Conventional Venous Sampling with Clinical Specimens for Direct Oral Anticoagulants Analysis by LC-MS/MS.Journal of mass spectrometry and advances in the clinical lab · 2026