Integrating Regulatory Planning and Metrology into Open-Source Medical Device Research: A Proposed Practice-Informed Framework from the A4IM Low-Field MRI Project.
Jitka Hanusova, Helena Kolarova, Jan Kavalirek, Pavel Korcak, Jiri Tesar
KAR BioTech s.r.o., náměstí Sítná 3105, 272 01 Kladno, Czech Republic. · Czech Metrology Institute, Okružní 772/31, 638 00 Brno, Czech Republic.
Key points
From the paper's conclusion- We derived a proposed regulatory-planning and metrology framework through a targeted multidisciplinary review, best-fit framework synthesis, and retrospective application to the A4IM low-field magnetic resonance imaging project.
Abstract
Open-source medical device research can accelerate technical progress, yet later adopters may need to reconstruct how each artefact relates to a system configuration, intended-use hypothesis, and evidence need. We derived a proposed regulatory-planning and metrology framework through a targeted multidisciplinary review, best-fit framework synthesis, and retrospective application to the A4IM low-field magnetic resonance imaging project. Binding requirements under Regulation (EU) 2017/745 were distinguished from guidance, voluntary standards, and research literature. The framework contains seven principles, eight functions, one canonical information model, and four reusable records. An analytical coverage audit mapped 42 source-derived relationship statements to the 8 functions; all were represented, while 5 overlaps were merged and 3 candidate relationships were split to preserve distinct responsibilities. Two public A4IM examples illustrate, but do not validate, the framework. For a 200 mm-diameter spherical volume, the magnet example distinguishes simulated design inhomogeneity (408 ppm), the as-built state (34,143 ppm), and the state after three shimming iterations (1602 ppm), with mean B0 48.47 mT at 24.7 °C. The acquisition example exposes which software and dataset identifiers are publicly recoverable and which remain unreported. The framework is intended to make configuration, risk, metrological evidence, provenance, and later adoption decisions traversable without treating research records as conformity evidence. External validation, implementation burden, time-to-market effects, and transferability remain prospective hypotheses.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- From Repeatability to Reproducibility in Quantitative Low-Field MRI: A Metrology-Informed Framework for Phantom Evidence.Journal of imaging · 2026
- The European Medical Device Ecosystem: Interactions Between Regulation, Innovation, Market Access, and Healthcare Systems.Healthcare (Basel, Switzerland) · 2026
- A Reference System for Open Source Portable Low-Field MRIarXiv · 2026
- Innovative Clinical Trial Approach for Evaluating Digital Medical Devices Under European Fast‐Track Regulatory FrameworksStatistics in Medicine · 2026
- Scaling medical device regulatory science using large language modelsnpj Digital Medicine · 2026
- Accuracy and reliability of privacy-compliant large language models for MRI protocol selection in radiological workflows.European radiology · 2026