Evaluation of Tendon Fastener Designed for Percutaneous Placement in an Ovine Model for Repair of Articular-Sided Rotator Cuff Tears.
Mark Miller, Adam Kohring, Tiffany Bridges, Benjamin C Gadomski, Kevin Labus, Daniel Regan, Jeremiah Easley, Drew W Koch,
Department of Orthopedic Surgery, Jefferson Health New Jersey, Stratford, NJ 08084, USA. · Department of Veterinary Medicine, Colorado State University College of Veterinary Medicine & Biomedical Sciences, Fort Collins, CO 80523, USA.
Key points
From the paper's conclusion- Conclusions: The fastener offers similar biomechanical stability to that of a screw-and-washer construct in the treatment of partial rotator cuff tears in an ovine model.
Abstract
Introduction: Surgical treatment for partial-thickness rotator cuff tendon (RCT) tears often involves open or arthroscopic conversion to a full-thickness tear. A novel method has been developed that utilizes sutureless bone-to-tendon fixation using a percutaneous fastener. Purpose: This study characterizes the biomechanical stability of the fastener in a partial-tear ovine rotator cuff model compared to that of a screw-and-washer. Methods: The study included an ex vivo cadaveric phase and an in vivo pilot. During the ex vivo phase, the infraspinatus tendon was partially transected in 20 cadaveric sheep shoulders. Shoulders were allocated into three groups: fastener (FS) (n = 10), screw-and-washer (S/W) (n = 7), and an empty group without fixation (n = 3). In an in vivo pilot, three sheep underwent bilateral partial infraspinatus transection and repair, with the fastener in one shoulder and the screw-and-washer in the contralateral shoulder; animals were euthanized at 8 weeks for histological analysis. Both cohorts underwent biomechanical assessments, and the in vivo samples underwent histological dissection and analysis. Results: No significant biomechanical differences were noted among the three groups in ultimate load (FS: 2822 N, S/W: 2599 N, empty: 2924 N), ultimate displacement (FS: 15.9 mm, S/W: 15.5 mm, empty: 13.8 mm), stiffness (FS: 252.4 N/mm, S/W: 237.7 N/mm, empty: 279.9 N/mm), cross-sectional area (FS: 90.0 mm2, S/W: 93.0 mm2, empty: 78.8 mm2), ultimate stress (FS: 32.9 MPa, S/W: 28.3 MPa, empty: 37.8 MPa), ultimate strain (FS: 0.203 mm/mm, S/W: 0.183 mm/mm, empty: 0.183 mm/mm), or elastic modulus (FS: 231.1 MPa, S/W: 220.3 MPa, empty: 271.3 MPa). The mode of failure was most frequently a mid-substance tear. Both the fastener and the screw-and-washer demonstrated comparable histology across the graded parameters regardless of the evaluated animal or sample. Conclusions: The fastener offers similar biomechanical stability to that of a screw-and-washer construct in the treatment of partial rotator cuff tears in an ovine model. Clinical Relevance: The results of this animal model study provide evidence to consider further clinical testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical Analysis of Rotator Cuff Repair: Comparison of Sutures and Fixation Techniques in a Synthetic Model.Cureus · 2026
- Surgical Technique for Arthroscopic Rotator Cuff Repair With All-Polyester Patch Augmentation.Arthroscopy techniques · 2026
- Arthroscopic Noninvasive Transtendon Double‐Pulley Suture‐Bridge Repair of Both‐Sided Partial‐Thickness Rotator Cuff TearsArthroscopy Techniques · 2026
- Arthroscopic Superior Capsular Reconstruction With Long-Head Biceps Tendon Rerouting Using Knotless Anchors.Arthroscopy techniques · 2026
- Comparison of an anchor plate with suture anchor and transosseous constructs for rotator cuff repair in a polyurethane foam block: a biomechanical studyBMC Musculoskeletal Disorders · 2026
- Round versus tape sutures with interference fit anchors perform similarly in cadaveric glenoid and greater tuberosity biomechanical testing.Archives of orthopaedic and trauma surgery · 2026