Round versus tape sutures with interference fit anchors perform similarly in cadaveric glenoid and greater tuberosity biomechanical testing.
James D Dimitroff, Conor Smith, Stefan Hanish, Lasun O Oladeji, Will Bezold, Clayton W Nuelle, Matthew J Smith
Department of Orthopaedic Surgery, University of Missouri, Columbia, USA. jdd7k9@umsystem.edu. · Orthopedic Surgery at the Mercy Center for Performance Medicine, Mercy Hospital, St Louis, USA.
Key points
From the paper's conclusion- Tape-type and round-type sutures demonstrated comparable biomechanical performance at the suture–anchor interface within the tested constructs.
Abstract
PURPOSE: To determine whether suture shape (flat-tape vs. round-wire) influences biomechanical performance at the suture-anchor-bone interface in interference-fit anchors used for rotator cuff and glenoid labrum repair. METHODS: Four matched pairs of human cadaveric shoulders were used for the glenoid model, with six anchor sites per glenoid. Five matched pairs of cadaveric humeri were used for the rotator cuff model, with three anchor sites per humerus. Each model tested multiple suture-anchor constructs under cyclic loading followed by load-to-failure. Biomechanical outcomes included cyclic displacement, yield characteristics, ultimate failure strength, stiffness, and force at fixed displacements. A linear mixed-effects model was used to compare suture types, with suture type and anchor position as fixed effects and specimen as a random effect. RESULTS: In the glenoid/labrum model, there were no significant differences among suture types in cyclic displacement (cycles 1, 50, or 100; p = 0.376, 0.101, 0.109), displacement at yield (p = 0.406), yield force (p = 0.880), force at 1 mm (p = 0.264) or 5 mm (p = 0.085), or ultimate failure strength (p = 0.204). Stiffness (p = 0.050) and force at 2 mm displacement (p = 0.033) differed at the overall model level, however, no significant pairwise differences were observed. In the humerus/rotator cuff model, displacement at cycle 50 (p = 0.022) and cycle 100 (p = 0.019) differed overall, although post hoc comparisons were not significant. Ultimate strength (p = 0.037), yield force (p = 0.012), and stiffness (p = 0.005) differed among groups, with post hoc analysis showing greater ultimate strength, yield force, and stiffness for 2-mm FiberTape compared with 2-mm XbraidTT. No other pairwise comparisons were significant. CONCLUSION: Suture shape did not meaningfully influence interference-fit fixation properties in the glenoid and humeral models evaluated in this study. Tape-type and round-type sutures demonstrated comparable biomechanical performance at the suture-anchor interface within the tested constructs. LEVEL OF EVIDENCE: Controlled Laboratory Study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of an anchor plate with suture anchor and transosseous constructs for rotator cuff repair in a polyurethane foam block: a biomechanical studyBMC Musculoskeletal Disorders · 2026
- Evaluation of Tendon Fastener Designed for Percutaneous Placement in an Ovine Model for Repair of Articular-Sided Rotator Cuff Tears.Journal of functional biomaterials · 2026
- Biomechanical Analysis of Rotator Cuff Repair: Comparison of Sutures and Fixation Techniques in a Synthetic Model.Cureus · 2026
- High-Tensile-Strength Suture Tape Provides Higher Load to Failure Than Conventional Suture for Glenoid Labral Repair in a Cadaveric Model.Arthroscopy, sports medicine, and rehabilitation · 2026
- Compression Suture Staple Repair of Simulated Gluteus Medius Partial Tendon Tears Provides Increased Contact Pressure Compared With Stepwise Bridged Repair.Arthroscopy, sports medicine, and rehabilitation · 2026
- Double-Cinch and Modified Mason-Allen Meniscal Root Repairs Demonstrate Comparable Time-Zero Biomechanical Performance with Cortical Button or Knotless Anchor Fixation: A Controlled Porcine Study.Journal of ISAKOS : joint disorders & orthopaedic sports medicine · 2026