Agreement and Trending Ability of the Non-Invasive Deep Body Temperature Sensor Moni-Patch™ in Critically Ill Patients Predominantly After Cardiovascular Surgery: A Single-Center Retrospective Comparison with Pulmonary Artery Blood Temperature
Masafumi Idei, Yohei Sakai, Yuri Maruyama, Shunsuke Takaki, Keiichiro Kawabata, Shinya Suganuma, Taichi Morimura, Shinya Oku,
Key points
From the paper's conclusion- Conclusions: Moni-Patch™ showed limited trending ability, and its error did not differ significantly from that of a zero-heat-flux thermometer.
Abstract
Background/Objectives: Continuous non-invasive core temperature monitoring may benefit intensive care unit (ICU) patients, but the performance of Moni-Patch™, a wireless sensor based on a compensated heat-flux method, has not been evaluated in critically ill patients. We assessed its agreement, trending ability, and fever detection against pulmonary artery blood temperature. Methods: In this single-center retrospective study, we analyzed one-minute data from adult ICU patients in whom pulmonary artery blood, bladder, zero-heat-flux (Bair Hugger™), and Moni-Patch™ temperatures were recorded simultaneously. Agreement was assessed using mixed-effects Bland–Altman analysis for repeated measures, with patient-level bootstrap confidence intervals. Trending ability (four-quadrant and polar analyses of 60 min changes) and fever detection (≥38.0 °C) were also evaluated. Results: Twenty-five patients contributed 14,705 paired measurements. Moni-Patch™ showed a bias of −0.30 °C (95% limits of agreement, −1.10 to +0.51 °C), with 72.2% of readings within ±0.5 °C. Its mean absolute error did not differ significantly from that of Bair Hugger™ (difference, −0.08 °C; 95% CI, −0.22 to 0.08), whereas bladder temperature agreed more closely (bias, +0.18 °C; limits, −0.22 to +0.58 °C). Directional concordance was 90.6% (77.3% on intervals defined by blood temperature changes). A reading ≥38.0 °C detected fever with a sensitivity of 36.6% and a specificity of 98.6%. Conclusions: Moni-Patch™ showed limited trending ability, and its error did not differ significantly from that of a zero-heat-flux thermometer. However, given its systematic underestimation and wide limits of agreement, caution is warranted when using it to replace invasive core temperature monitoring or as the sole means of detecting fever.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study.Sensors (Basel, Switzerland) · 2026
- A Comprehensive Review of Non-Invasive Core Body Temperature Measurement TechniquesSensors · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Clinical applicable skin-interfaced thermal guiding sensor for warning perioperative abnormalities in core body temperature and blood perfusion rate.Nature communications · 2026
- Dynamic Infrared Thermography for Quantitative Assessment of Peripheral Microcirculation: Models, Metrics, and Clinical TranslationPhysiological Measurement · 2026
- Esophageal thermometry outperforms bladder and tympanic monitoring during targeted temperature management and shock in the ICU.Journal of critical care · 2026