Agreement and Trending Ability of the Non-Invasive Deep Body Temperature Sensor Moni-Patch™ in Critically Ill Patients Predominantly After Cardiovascular Surgery: A Single-Center Retrospective Comparison with Pulmonary Artery Blood Temperature

Masafumi Idei, Yohei Sakai, Yuri Maruyama, Shunsuke Takaki, Keiichiro Kawabata, Shinya Suganuma, Taichi Morimura, Shinya Oku,

Observational

Key points

From the paper's conclusion
  • Conclusions: Moni-Patch™ showed limited trending ability, and its error did not differ significantly from that of a zero-heat-flux thermometer.

Abstract

Background/Objectives: Continuous non-invasive core temperature monitoring may benefit intensive care unit (ICU) patients, but the performance of Moni-Patch™, a wireless sensor based on a compensated heat-flux method, has not been evaluated in critically ill patients. We assessed its agreement, trending ability, and fever detection against pulmonary artery blood temperature. Methods: In this single-center retrospective study, we analyzed one-minute data from adult ICU patients in whom pulmonary artery blood, bladder, zero-heat-flux (Bair Hugger™), and Moni-Patch™ temperatures were recorded simultaneously. Agreement was assessed using mixed-effects Bland–Altman analysis for repeated measures, with patient-level bootstrap confidence intervals. Trending ability (four-quadrant and polar analyses of 60 min changes) and fever detection (≥38.0 °C) were also evaluated. Results: Twenty-five patients contributed 14,705 paired measurements. Moni-Patch™ showed a bias of −0.30 °C (95% limits of agreement, −1.10 to +0.51 °C), with 72.2% of readings within ±0.5 °C. Its mean absolute error did not differ significantly from that of Bair Hugger™ (difference, −0.08 °C; 95% CI, −0.22 to 0.08), whereas bladder temperature agreed more closely (bias, +0.18 °C; limits, −0.22 to +0.58 °C). Directional concordance was 90.6% (77.3% on intervals defined by blood temperature changes). A reading ≥38.0 °C detected fever with a sensitivity of 36.6% and a specificity of 98.6%. Conclusions: Moni-Patch™ showed limited trending ability, and its error did not differ significantly from that of a zero-heat-flux thermometer. However, given its systematic underestimation and wide limits of agreement, caution is warranted when using it to replace invasive core temperature monitoring or as the sole means of detecting fever.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers