Mapping Efficiency of an Octaspline Versus Pentaspline Multielectrode Catheter for High-Density Mapping of Regular Atrial Tachyarrhythmias: A Multicenter Randomized Cross-Over Trial.
Bakhtawar K Mahmoodi, Jippe C Balt, Muchtiar Khan, Maurits C E F Wijffels, Gijsbert S de Ruiter, Tamas Szili-Torok, Sing-Chien Yap
Department of Cardiology, Thorax Center, Cardiovascular Institute, Erasmus Medical Center, 3015 GD Rotterdam, The Netherlands. · Department of Cardiology, St. Antonius Hospital, 3435 CM Nieuwegein, The Netherlands.
Key points
From the paper's conclusion- We compared the mapping efficiency of the Octaray and Pentaray catheters in patients undergoing AT mapping.
Abstract
Background: Catheter ablation of regular atrial tachyarrhythmias (ATs) relies on high-density electroanatomical mapping and can be technically challenging. We compared the mapping efficiency of the Octaray and Pentaray catheters in patients undergoing AT mapping. Methods: Patients with either left or right AT underwent three-dimensional Coherent mapping using both the Octaray and Pentaray catheters in a randomized cross-over design. The primary endpoint was the total number of electrograms acquired per map at a predefined filling threshold of 5 mm. Results: A total of 46 patients with left or right AT were enrolled across three ablation centers. Successful mapping with both catheters was achieved in 39 patients (20 left ATs and 19 right ATs). The median number of electrograms acquired per map (7437 vs. 4679; p < 0.001) and median electrogram density (50.2 vs. 29.8 electrograms/cm2; p < 0.001) was significantly higher with Octaray than with Pentaray. Furthermore, median mapping times were significantly shorter with Octaray (5.4 vs. 8.1 min; p < 0.001), resulting in an approximately 2.5-fold higher acquisition rate (1489 vs. 602 electrograms/minute; p < 0.001). In addition, Octaray demonstrated a significantly lower proportion of internal points compared with Pentaray, particularly in low-voltage regions (bipolar voltage < 0.5 mV; 22% vs. 44%; p = 0.01). Conclusions: Compared with Pentaray, use of the Octaray catheter facilitated faster high-density electroanatomical mapping of complex ATs, thereby improving mapping efficiency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Analysis of Multielectrode Mapping Catheter Designs in Scar-Related Ventricular Tachycardia Ablation.Pacing and clinical electrophysiology : PACE · 2026
- Sparse-EWI: Towards an automated and real time focal arrhythmia mapping technique - in silico to clinical validation.IEEE transactions on ultrasonics · 2026
- Matching Electrode Configuration to Signal Sensing for Cardiac Electroanatomic Mapping: A Critical Appraisal.Indian pacing and electrophysiology journal · 2026
- One-year outcomes in patients with persistent atrial fibrillation using CARTOFINDER mapping with OCTARAY: a Shinshu-AB registry study.Indian pacing and electrophysiology journal · 2026
- Calibration and processing sensitivity of time-embedded three-dimensional electrocardiographic descriptors: a two-dataset benchmarkmedRxiv · 2026
- Pro-arrhythmogenicity of single-tip vs. multi-spline catheters in post-MI VT ablation: a prospective, international, two-center experiencemedRxiv · 2026