Comparative Analysis of Multielectrode Mapping Catheter Designs in Scar-Related Ventricular Tachycardia Ablation.
Nadica Trajkovska, Moneeb Khalaph, Mustapha El Hamriti, Philipp Lucas, Martin Braun, Stephan Molatta, Thomas Fink, Vanessa Sciacca,
Department of Electrophysiology, Heart and Diabetes Center NRW, Ruhr University Bochum, Bad Oeynhausen, Germany. · Department For Cardiology/Electrophysiology, GZO-Spital Wetzikon, Wetzikon, Zurich, Switzerland.
Key points
From the paper's conclusion- Both catheter designs achieved comparable acute procedural success, procedural efficiency, and safety in scar-related VT ablation. Procedures with the spline catheter were associated with a higher inducible VT morphology burden; this hypothesis-generating finding, which may partly reflect differences in substrate location, warrants prospective evaluation.
Abstract
BACKGROUND: High-density (HD) electroanatomical mapping is routinely used for scar-related ventricular tachycardia (VT) ablation. The performance of multielectrode mapping catheters with different array designs has not been directly compared. OBJECTIVE: To compare the procedural performance of two different multielectrode mapping catheter designs-a flexible spline catheter (SC) and a fixed paddle catheter (PC) in scar-related VT ablation. METHODS: We retrospectively analyzed 104 consecutive patients who underwent left ventricular scar-related VT ablation (SC, n = 44; PC, n = 60). The primary endpoint was acute procedural success, defined as non-inducibility of the clinical VT after ablation. Secondary endpoints included procedural efficiency, VT inducibility, VT morphology burden, and procedural and in-hospital adverse events. RESULTS: Acute procedural success was achieved in 95.5% of SC and 98.3% of PC procedures (p = 0.57). Procedure duration, LV mapping time, mapping point count, fluoroscopy time, and adverse-event rates were comparable between groups. Overall VT inducibility (mechanical or programmed induction) was similar (81.8% vs. 76.7%; p = 0.63); however, ≥3 distinct inducible VT morphologies were significantly more frequent with the spline catheter than with the paddle catheter (38.6% vs. 11.7%; p = 0.002). During follow-up, rates of VT recurrence (34.5% vs. 20.0%; p = 0.19) and repeat ablation (13.8% vs. 7.3%; p = 0.44) did not differ significantly between groups. CONCLUSIONS: Both catheter designs achieved comparable acute procedural success, procedural efficiency, and safety in scar-related VT ablation. Procedures with the spline catheter were associated with a higher inducible VT morphology burden; this hypothesis-generating finding, which may partly reflect differences in substrate location, warrants prospective evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pro-arrhythmogenicity of single-tip vs. multi-spline catheters in post-MI VT ablation: a prospective, international, two-center experiencemedRxiv · 2026
- Scar-related ventricular tachycardia ablation using a dual-energy lattice-tip ablation system guided by pre-procedural imagingEuropean Heart Journal · 2026
- Feasibility and Safety of Strategic Multielectrode Positioning Guided by S3 Protocol for Hemodynamically Unstable Ventricular Tachycardia: The S3-STaMP Project.JACC. Clinical electrophysiology · 2026
- Mapping Efficiency of an Octaspline Versus Pentaspline Multielectrode Catheter for High-Density Mapping of Regular Atrial Tachyarrhythmias: A Multicenter Randomized Cross-Over Trial.Journal of cardiovascular development and disease · 2026
- Computed tomography-guided vs conventional catheter ablation for ventricular tachycardia: the InEurHeart trialEuropean Heart Journal · 2026
- Three-Dimensional Perspective on Scar-Related Ventricular Tachycardia Substrate: Critical Conduction Abnormalities as Ablation Targets and Implications for Procedural Endpoints.Reviews in cardiovascular medicine · 2026