Effect of LED Light-Curing Distance on Microhardness and Bracket Bonding Performance: An In Vitro Study.
Brissa Fuentes-Ortiz, Roberto Ruiz-Diaz, Benjamin Aranda-Herrera, Jose Vivaldi, Silvia Rojas-Rueda, Francisco X Azpiazu-Flores, Rogelio Jose Scougall-Vilchis, Carlos A Jurado,
Orthodontics Area, National School of Higher Studies (ENES) León Unit, National Autonomous University of Mexico (UNAM), León 37689, Guanajuato, Mexico. · School of Dentistry, Meritorious Autonomous University of Puebla (BUAP), Puebla 72530, Puebla, Mexico.
Key points
From the paper's conclusion- This study evaluated the effect of 0-, 5-, and 10-mm LED curing distances on resin microhardness, orthodontic bracket shear bond strength, and adhesive remnant index (ARI).
Abstract
Objectives: The distance between a light-curing tip and an orthodontic adhesive may alter the energy reaching the resin and affect polymerization-related mechanical performance. This study evaluated the effect of 0-, 5-, and 10-mm LED curing distances on resin microhardness, orthodontic bracket shear bond strength, and adhesive remnant index (ARI). Methods: Orthodontic resin was photoactivated for 10 s at 0, 5, or 10 mm. Vickers microhardness was measured in 90 independent resin specimens (30/group; three indentations/specimen) using a 5-N load for 5 s. In a separate tooth-based experiment, 45 extracted permanent human teeth received conventional metallic brackets. After 24 h of water storage, shear bond strength was tested at 1 mm/min and ARI was assessed microscopically. Results: Microhardness decreased from 50.01 ± 11.47 HV at 0 mm to 30.24 ± 7.49 HV at 5 mm and 25.03 ± 8.37 HV at 10 mm (p < 0.001). The difference between 5 and 10 mm was not statistically significant (p = 0.081). Shear bond strength decreased from 4.76 ± 1.31 MPa to 2.46 ± 0.92 MPa and 0.95 ± 0.69 MPa, respectively (p < 0.001). High ARI scores predominated, and the four-category ARI analysis did not detect a significant difference among groups (p = 0.270). Conclusions: Increasing LED curing-tip distance was associated with lower immediate microhardness and bracket bond strength. Microhardness was significantly lower at 5 and 10 mm than at 0 mm, whereas bond strength decreased significantly across all three distances. These outcomes were obtained in separate experimental systems and were interpreted as complementary rather than mechanistically linked. Clinically, the curing tip should be positioned as close and perpendicular to the bracket as access permits, while recognizing that these in vitro findings do not predict long-term bracket survival.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of light-emitting diode curing tip distance on shear bond strength in metal, ceramic and polycarbonate orthodontic brackets: An in vitro study.Bioinformation · 2026
- The effect of different intensity modes and curing times of new-generation LED light-curing units and orthodontic adhesives on shear bond strength of metal brackets.Korean journal of orthodontics · 2026
- Effect of high irradiance and short exposure times on the shear bond strength of metal and ceramic brackets.BMC oral health · 2026
- In vitro evaluation of the efficacy of a new orthodontic primer in metallic brackets cementationClinical Oral Investigations · 2026
- Effect of Light-Curing Distance on Polymerization Behavior of Resin Composites: A Systematic Review and Meta-AnalysisPhyschem · 2026
- Effect of different lengths of post-curing UV light exposure on the mechanical properties of 3D-printed aligners: An in vitro studyJournal of Orthodontics · 2025