Dimensional Stability of SLA 3D-Printed Surgical Guide Resin After Steam Sterilization (121 °C and 134 °C) and Isopropyl Alcohol Disinfection: An In Vitro Study
Ioan Sîrbu, Andreea Mihaela Custura, Adelin Radu, Vlad Gabriel Vasilescu, Vladimir Năstasie, VASILE IULIAN ANTONIAC, Marian Miculescu, Adrian Ionuţ Şişman,
Carol Davila University of Medicine and Pharmacy · Universitatea Națională de Știință și Tehnologie Politehnica București
Key points
From the paper's conclusion- Objectives: This in vitro study evaluated the dimensional changes of SLA 3D-printed Surgical Guide Resin V1 (Formlabs) after steam sterilization at 121 °C (AUT121) and 134 °C (AUT134) and after disinfection using 70% isopropyl alcohol (IPA70), compared with an untreated control group.
Abstract
Background: Additively manufactured surgical guides require post-processing and subsequent decontamination prior to intraoral use. Steam sterilization and chemical disinfection protocols may influence the dimensional stability of polymer-based guide materials and potentially affect clinical fit and accuracy. Objectives: This in vitro study evaluated the dimensional changes of SLA 3D-printed Surgical Guide Resin V1 (Formlabs) after steam sterilization at 121 °C (AUT121) and 134 °C (AUT134) and after disinfection using 70% isopropyl alcohol (IPA70), compared with an untreated control group. Methods: Forty standardized specimens were fabricated using SLA technology and divided into four groups (n = 10/group): Control (CT), 121 °C steam sterilization (AUT121), 134 °C steam sterilization (AUT134), and IPA70 disinfection. Two linear measurement zones (L1 and L2) were assessed per specimen. Baseline measurements were recorded with a caliper (mm). Post-treatment measurements were obtained using microscopic evaluation, recorded in µm, and converted to mm for analysis. Dimensional change was calculated as ΔL = L_after − L_before. Within-group comparisons and between-group analyses were performed with a significance level of α = 0.05. Results: Steam sterilization at 134 °C (AUT 134) produced statistically significant dimensional changes in both zones (L1: p = 0.036; L2: p = 0.042). No statistically significant differences were observed in the AUT121 group (L1: p = 0.437; L2: p = 0.682) or the IPA70 group (L1: p = 0.164; L2: p = 0.086). Between-group analysis showed no significant differences for ΔL1 (p = 0.345), whereas ΔL2 differed significantly among groups (p = 0.021). Conclusions: Under the conditions of this study, AUT134 steam sterilization significantly affected the dimensional stability of SLA-printed Surgical Guide Resin V1 specimens. The AUT121 protocol and IPA70 disinfection did not result in statistically significant dimensional changes compared with baseline.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dimensional Accuracy of 3D-Printed Surgical Guides in Relation to Design Modifications and Fabrication Workflow: An Exploratory Laboratory Pilot StudyDentistry Journal · 2026
- Sterilization Effect on Axes and Distances Between Three Implants in Curved Lines When Using Two Types of Two-Piece Intra-Oral Scan Bodies: An In Vitro Study.Dentistry journal · 2026
- Trueness and stability of additively manufactured removable dies: An in vitro study on plant-based resins and alcohol-free cleaning solutionsJournal of Prosthetic Dentistry · 2026
- Effect of disinfection protocols on the dimensional stability of commercial vinyl polyether silicone impression materials: a systematic review and meta-analysis.BMC oral health · 2026
- Construction and demolition waste-reinforced photopolymer resin for stereolithography (SLA) 3D printingJournal of Materials Science: Materials in Engineering · 2026
- Comparative Evaluation of SLA and DLP 3D Printing in Dental Implant Guides: Impact on Fabrication Accuracy, Speed, and Resin UsageDentistry Journal · 2025