Dimensional Stability of SLA 3D-Printed Surgical Guide Resin After Steam Sterilization (121 °C and 134 °C) and Isopropyl Alcohol Disinfection: An In Vitro Study

Ioan Sîrbu, Andreea Mihaela Custura, Adelin Radu, Vlad Gabriel Vasilescu, Vladimir Năstasie, VASILE IULIAN ANTONIAC, Marian Miculescu, Adrian Ionuţ Şişman,

Carol Davila University of Medicine and Pharmacy · Universitatea Națională de Știință și Tehnologie Politehnica București

Preclinical · n=10

Key points

From the paper's conclusion
  • Objectives: This in vitro study evaluated the dimensional changes of SLA 3D-printed Surgical Guide Resin V1 (Formlabs) after steam sterilization at 121 °C (AUT121) and 134 °C (AUT134) and after disinfection using 70% isopropyl alcohol (IPA70), compared with an untreated control group.

Abstract

Background: Additively manufactured surgical guides require post-processing and subsequent decontamination prior to intraoral use. Steam sterilization and chemical disinfection protocols may influence the dimensional stability of polymer-based guide materials and potentially affect clinical fit and accuracy. Objectives: This in vitro study evaluated the dimensional changes of SLA 3D-printed Surgical Guide Resin V1 (Formlabs) after steam sterilization at 121 °C (AUT121) and 134 °C (AUT134) and after disinfection using 70% isopropyl alcohol (IPA70), compared with an untreated control group. Methods: Forty standardized specimens were fabricated using SLA technology and divided into four groups (n = 10/group): Control (CT), 121 °C steam sterilization (AUT121), 134 °C steam sterilization (AUT134), and IPA70 disinfection. Two linear measurement zones (L1 and L2) were assessed per specimen. Baseline measurements were recorded with a caliper (mm). Post-treatment measurements were obtained using microscopic evaluation, recorded in µm, and converted to mm for analysis. Dimensional change was calculated as ΔL = L_after − L_before. Within-group comparisons and between-group analyses were performed with a significance level of α = 0.05. Results: Steam sterilization at 134 °C (AUT 134) produced statistically significant dimensional changes in both zones (L1: p = 0.036; L2: p = 0.042). No statistically significant differences were observed in the AUT121 group (L1: p = 0.437; L2: p = 0.682) or the IPA70 group (L1: p = 0.164; L2: p = 0.086). Between-group analysis showed no significant differences for ΔL1 (p = 0.345), whereas ΔL2 differed significantly among groups (p = 0.021). Conclusions: Under the conditions of this study, AUT134 steam sterilization significantly affected the dimensional stability of SLA-printed Surgical Guide Resin V1 specimens. The AUT121 protocol and IPA70 disinfection did not result in statistically significant dimensional changes compared with baseline.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Dimensional Stability of SLA 3D-Printed Surgical Guide Resin After Steam Sterilization (121 °C and 134 °C) and Isopropyl Alcohol Disinfection: An In Vitro Study — Science Skim