Evaluation of a new non-competitive chemiluminescent immunoassay for 25-OH vitamin D measurement.
Jinqing Zhong, Binning Huang, Aihua Lin, Geng Wu, Lili Zhang, Xiaowei Yu, Shun Zhuang, Daying Zheng,
Department of Laboratory Medicine, Xiamen Humanity Hospital, Xiamen, China. · Department of Laboratory Medicine, Xiamen Humanity Maternity Hospital, Xiamen, China.
Key points
From the paper's conclusion- The Snibe-NACA 25-OH Vitamin D assay demonstrated good performance and compared favorably with both LC-MS/MS and other competitive assays, providing evidence of its high reliability in clinical diagnostic practice.
Abstract
OBJECTIVES: The increasing demand for vitamin D measurement over the past few decades has prompted manufacturers to develop automated chemiluminescence immunoassay (CLIA), including the noncompetitive anti-immune complex assay (NACA) system for the measurement of small molecules. This study aims to evaluate the performance of the Snibe-NACA 25-OH Vitamin D assay for measuring total 25-Hydroxy Vitamin D assay [25(OH)D] levels. METHODS: A total of 200 serum samples were collected from a clinical laboratory for 25(OH)D assay to evaluate the performance of the Snibe-NACA 25-OH Vitamin D assay using a double-blind method. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) was used as a reference method and the results were compared with those obtained from four other chemiluminescence competitive 25(OH)D assays. RESULTS: The Snibe-NACA 25-OH Vitamin D assay demonstrated satisfactory analytical performance with optimal precision (repeatability CV: 1.13-2.75 %;intra-laboratory CV: 1.59-2.58 %) and relatively small bias (-3.84 % mean relative bias compared with the LC-MS/MS),excellent agreement with LC-MS/MS, as evidenced by Passing-Bablok regression (R2: 0.9245, slope: 0.99, 95 % CI: 0.96-1.03; intercept: -1.24, 95 % CI: -3.49 to 0.65) and high correlation (Spearman's ρ=0.969; ICC=0.9584). Notably, the assay showed the highest concordance with LC-MS/MS in the classification of vitamin D status, achieving an overall agreement of 87.0 %, with particularly strong concordance in identifying deficiency (97.8 %). CONCLUSIONS: The Snibe-NACA 25-OH Vitamin D assay demonstrated good performance and compared favorably with both LC-MS/MS and other competitive assays, providing evidence of its high reliability in clinical diagnostic practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of LC-MS/MS and CLIA Methods for Vitamin D Measurement in an Unselected Outpatient Cohort: Statistical Evaluation and Impact on Clinical Classification.Diagnostics (Basel, Switzerland) · 2026
- Analytical performance of a single epitope meta-sandwich total 25-hydroxyvitamin D immunoassay and establishment of reference intervals in a Chinese population.Annals of clinical biochemistry · 2026
- Performance evaluation of a direct chemiluminescence assay for 25-hydroxyvitamin D and its application in assessing vitamin D status among reproductive-age women in southeastern China.PloS one · 2026
- Comprehensive Comparison of Three Automated 25-Hydroxyvitamin D Immunoassays with a VDSCP Traceable LC-MS/MS Method Across Clinically Relevant Decision Thresholds.Nutrients · 2026
- Is a low-cost latex immunoturbidimetric assay suitable for routine 25(OH)D screening? Analytical comparison with chemiluminescent immunoassay and LC-MS/MS.Scandinavian journal of clinical and laboratory investigation · 2026
- B-029 Analytical Performance Evaluation of the Double-Antibody Sandwich Mindray Total-Vitamin D Assay and Comparison Versus Liquid Chromatography Tandem Mass Spectrometry and Commercial ImmunoassaysClinical Chemistry · 2026