Chemosensors

Sensitive Determination of Iron(II) Cations in Pharmaceutical Products Using a Modified Gold Thin-Layer Screen-Printed Carbon Electrode

Irina Mirela Apetrei, Constantin Apetrei

Key points

From the paper's conclusion
  • The results were satisfactory, with recoveries ranging from 99.1% to 103.3% and values close to those reported by the producers.

Abstract

In this study, a new electrochemical sensor was fabricated by depositing a gold (Au) layer onto a screen-printed carbon electrode (SPCE) via chronoamperometry, followed by a heat treatment process. The morphology of the gold-modified electrode (Au-SPCE) was examined using scanning electron microscopy, revealing a homogeneous metal layer alongside the presence of gold microcrystals. Preliminary studies in a KNO3 supporting electrolyte, ferrocyanide/ferricyanide, and Fe(NO3)2 solutions demonstrated that the electrode possesses favorable electrochemical properties, such as high sensitivity, rapid electron transfer, and adequate stability. The Au-SPCE sensor was tested in Fe(NO3)2 solutions in a concentration range of 1–300 nM using square wave voltammetry and chronoamperometry (at an applied potential of 0.52 V), yielding a broad linear range (1–100 nM). The calculated limits of detection were 0.50 nM (for square wave voltammetry) and 0.18 nM (for chronoamperometry)—values superior to those reported for other sensors in the literature. The Au-SPCE sensor was also evaluated for the quantitative determination of Fe(II) in various pharmaceutical formulations using both the standard-addition method and the calibration curve interpolation method. The results were satisfactory, with recoveries ranging from 99.1% to 103.3% and values close to those reported by the producers. The development of this sensor offers an effective approach for the quality control of pharmaceutical products containing Fe(II).

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Sensitive Determination of Iron(II) Cations in Pharmaceutical Products Using a Modified Gold Thin-Layer Screen-Printed Carbon Electrode — Science Skim