Development of a Calibration Transfer Methodology and Experimental Setup for Urine Headspace Analysis
Michela Cassinerio, Beatrice Julia Lotesoriere, Stefano Robbiani, Emanuele Zanni, Fabio Grizzi, Gianluigi Taverna, Raffaele L. Dellacá, Laura Capelli
Politecnico di Milano · IRCCS Humanitas Research Hospital
Key points
From the paper's conclusion- These results demonstrate that combining reproducible synthetic standards with DS enables effective model transfer across E-Noses, reducing calibration requirements and supporting their broader applicability in medical diagnostics.
Abstract
Electronic noses (E-Noses) equipped with metal-oxide semiconductor (MOS) sensors are promising tools for non-invasive medical diagnostics. Their adoption in clinical practice, however, is limited—among others—by sensor variability across devices, which makes individual calibration necessary. This study presents an approach for the development of a calibration transfer (CT) methodology for urine headspace analysis, involving the design and realization of a dedicated experimental setup and protocol. Partial least squares-discriminant analysis (PLS-DA) models were trained on human urine samples enriched with selected biomarkers to simulate pathological states. Models from a reference (“master”) device were transferred to other (“slave”) units in multiple master–slave configurations using Direct Standardization (DS). To overcome the variability of human urine, synthetic urine recipes were formulated to mimic sensor responses and serve as reproducible transfer samples. Several strategies for selecting transfer samples were evaluated, including the Kennard–Stone algorithm, a DBSCAN-based approach, and random selection. Without CT, classification accuracy on slave devices decreased markedly (37–55%) compared to the master’s performance (79%), whereas applying DS with synthetic standards restored accuracy to 75–80%. These results demonstrate that combining reproducible synthetic standards with DS enables effective model transfer across E-Noses, reducing calibration requirements and supporting their broader applicability in medical diagnostics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Isolating the Contributions of Resistance-Ratio Normalization and Humidity Fusion in a Minimal Three-Sensor Electronic Nose.Sensors (Basel, Switzerland) · 2026
- Disentangling Response Magnitude and Relative MOS Sensor Fingerprints for Electronic-Nose Screening of Off-Odor in Table Olives.Sensors (Basel, Switzerland) · 2026
- Deep Learning-Enabled Interpretative SERS Analysis of Key Urinary Metabolites in Nephropathy Using Interfacial Charge-Engineered Silver Nanoparticles.Analytical chemistry · 2026
- A low-cost room-temperature electronic nose prototype for gas sensor characterization and simulated breath analysisMeasurement Science and Technology · 2026
- Study of bladder cancer detection in standard white light versus AI-supported endoscopy-01 (RAISE-01) – Development and validation of an AI-based support toolInternational Journal of Medical Informatics · 2026
- Overcoming Cross-Sensitivity for the Accurate Identification of Acetone, Isopropanol, and Clinical Mixtures Using a MEMS Dual-Sensor Array and Multi-task Deep Learning.ACS sensors · 2026