Combined Use of Savi Scout and Magtrace for Non-Palpable Breast Cancer: A Retrospective Study.
Claire Lloyd-Davies, Abdul Kasem
London North West University Healthcare NHS Trust, London HA1 3UJ, UK. · King's College Hospital NHS Foundation Trust, London BR6 8ND, UK.
Key points
From the paper's conclusion- This study evaluates the combination of Savi Scout and Magtrace and demonstrates a practical approach for simultaneous tumor localization and SLN identification.
Abstract
BACKGROUND: Non-palpable breast cancer management has become less invasive and more efficient, driven by advances in tumor localization and sentinel lymph node (SLN) identification. Savi Scout, using a radar-based localization system, can be implanted into the tumor before neoadjuvant systemic therapy (NST). Magtrace, a superparamagnetic iron oxide nanoparticle for SLN detection, can be administered preoperatively. This study evaluated combining Savi Scout and Magtrace for breast-conserving surgery and SLN biopsy in non-palpable breast cancer. METHODS: Patients who underwent breast-conserving surgery and SLN biopsy for non-palpable breast cancer using Savi Scout and Magtrace between December 2022 and June 2026 at HCA Healthcare London Bridge Hospital were retrospectively identified. Outcomes included successful tumor localization, SLN identification, intraoperative signal detection failure, margin status, re-excision rate and staining. RESULTS: 72 patients were identified (mean age 52 years, standard deviation ± 10), with a total of 78 breast lesions and 73 SLN biopsies. A total of 27.8% (20/72) underwent neo-adjuvant therapy with Savi Scout in situ. SLNs were identified using Magtrace in 95.9% (70/73) of SLN biopsies, with a mean of 2.7 (standard deviation ± 0.7) SLNs identified per procedure. A total of 78 Savi Scout devices were implanted, one for each lesion. Of these, 97.4% (76/78) were localizable intraoperatively, 100.0% (78/78) were retrieved and none migrated. One patient required re-excision for margin involvement. CONCLUSIONS: Combined use of Savi Scout and Magtrace is feasible, demonstrates high technical success, and offers pre-operative imaging flexibility in primary surgery for non-palpable breast cancers and post-NST patients. This study evaluates the combination of Savi Scout and Magtrace and demonstrates a practical approach for simultaneous tumor localization and SLN identification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simultaneous Use of Radar and Magnetic Navigation in Breast Cancer Surgery.Cancers · 2026
- A Multicenter Non‐Inferiority Randomized Controlled Trial of Sentinel Lymph Node Biopsy in Early Breast Cancer: Conventional Dual Tracer Versus Superparamagnetic Iron OxideWorld Journal of Surgery · 2026
- Robotic sentinel lymph node biopsy with dual-tracer mapping in 76 patients with breast cancer: high detection rate and low false-negative rate.Journal of robotic surgery · 2026
- SCOUT Radar Localization System-Guided Excision of Nonpalpable Breast Lesions: Initial Experience in a Latin American Population.The breast journal · 2026
- A Prospective Randomized Trial Comparing Radioguided Occult Lesion Localization (ROLL) and Magnetic Seed Localization for the Localization of Nonpalpable Breast Lesions: Analysis of Surgical Outcomes, Patient’s Perception, and CostsAnnals of Surgical Oncology · 2026
- Comparative analysis of wire-guided localization versus Pintuition magnetic marker system in non-palpable breast cancer surgery: a single-center, retrospective cohort study.Annals of medicine and surgery (2012) · 2026