A Wired Pressure-Sensing Smart Biliary Stent: Bench Characterization and Feasibility in a Porcine Model
Jonghyun Lee, Sung Yong Han, Seung Min Hong, Dong Chan Joo, Kiyoun Yi, Jae Won Lee, Young Woo Kwon
Key points
From the paper's conclusion- We developed a pressure-sensing smart biliary stent by integrating a flexible contact-resistive sensor, comprising opposing Au mesh electrodes on liquid crystal polymer films, into a commercial self-expandable metal stent connected to an external monitoring system.
Abstract
Biliary stents restore drainage in malignant biliary obstruction, but recurrent dysfunction creates a need to monitor the patency of the stent. We developed a pressure-sensing smart biliary stent by integrating a flexible contact-resistive sensor, comprising opposing Au mesh electrodes on liquid crystal polymer films, into a commercial self-expandable metal stent connected to an external monitoring system. Under loading through a compliant pad matched to the 1 mm × 2 mm sensing window (n = 4 sensors), the normalized resistance decreased approximately linearly up to 2.5 kPa and approached a plateau at approximately 3.0 kPa, with a sensitivity of −37.0 ± 0.4% kPa−1 in the linear region, a pressure hysteresis of 4.1 ± 1.5% of full scale, and, in a representative sensor, a variation within ±2.1% of the mean over five loading–unloading cycles. Between 30 and 45 °C, measured in the solution at the stent, the step means of ΔR/R0 remained within 0.10% of zero (n = 4 sensors). In recirculating model bile inoculated with Escherichia coli, the sensor output in the run with a complete recording began to change after approximately 400 min and reached the plateau of the response after approximately 840 min, whereas in sterile model bile the mean output of four sensors remained within 0.15% of baseline over 14 h. In a porcine biliary stricture model, the device was deployed endoscopically in one animal, its wires were externalized through a percutaneous endoscopic gastrostomy tube, and the monitor, zeroed in vivo after connection, displayed a change in readout after start-up, which was not verified against a reference pressure. The connection wire broke inside the percutaneous endoscopic gastrostomy tube within 1 week and had also broken at several steps of the procedure when it was pulled; in bench tests, it fractured at a monotonic tensile load of 1.878 N. As an initial feasibility study toward clinical application, these results indicate that sensor integration, endoscopic delivery, wire routing, and external readout can be achieved, identify the procedural steps at which the sensor and its wires failed, and define the extension of the measurement range and the protection of the connection that are required before occlusion monitoring can be evaluated in vivo.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A PTBD-Based Smart Biliary Stent for Continuous Intrabiliary Pressure Monitoring: Development and In Vivo FeasibilityBiosensors · 2026
- Biostable wireless sensor-integrated bioresorbable stent for real-time monitoring of vascular pressure and fractional flow reserveMicrosystems & Nanoengineering · 2026
- Smart Stents as Wireless Sensing Platforms: A Subsystem-Level ReviewMicromachines · 2026
- Bioactive shape-memory biodegradable biliary stent functionalized with dHAM-Curcumin for improved biliary repair and microenvironment modulation.Biomaterials · 2026
- Injectable Bioadhesive Ultrasonic Sensor for Continuous and Wireless On-Demand Pressure Monitoring in Acute Compartment Syndrome.Advanced materials (Deerfield Beach, Fla.) · 2026
- A Soft, Conformal, and Implantable Sensor for On-Demand Bladder Volume Monitoring.ACS applied bio materials · 2026