A PTBD-Based Smart Biliary Stent for Continuous Intrabiliary Pressure Monitoring: Development and In Vivo Feasibility
Jonghyun Lee, Young Woo Kwon, Hye Ji Lee, Sung Hee Park, Hyung Il Seo, Suk Kim, Seung Baek Hong, Sung Yong Han,
Key points
From the paper's conclusion- These findings provide preliminary evidence for the technical feasibility of PTBD-based smart biliary stent monitoring and suggest that intrabiliary pressure may serve as a real-time physiological indicator of stent dysfunction.
Abstract
Biliary stent dysfunction in malignant biliary obstruction remains clinically occult until jaundice or cholangitis develops because current surveillance methods yield no continuous functional information. This study presents a percutaneous transhepatic biliary drainage (PTBD)-based smart biliary stent incorporating a thin-film pressure sensor and evaluates its in vivo feasibility for continuous intrabiliary pressure monitoring. A resistive pressure sensor was fabricated on a liquid-crystal polymer substrate with gold-mesh electrodes encapsulated for bile exposure and integrated into a covered self-expandable metal stent. Following in vitro characterization under bile flow conditions, the device was evaluated in a porcine biliary stricture model created using endoscopic radiofrequency ablation. The sensor lead wire was externalized through a mature PTBD tract. The sensor exhibits a pressure-dependent resistance response and maintains electrical continuity after stent compression and expansion. Of the three animals, one completed the full workflow, from stricture formation to lead-wire externalization, allowing stable continuous pressure recording for approximately 48 h; the in vivo evaluation was therefore based on a single animal. After endoscopic hemoclip-induced distal stent occlusion, the intrabiliary pressure increased from approximately 200 Pa to approximately 3800 Pa within approximately 25 min. These findings provide preliminary evidence for the technical feasibility of PTBD-based smart biliary stent monitoring and suggest that intrabiliary pressure may serve as a real-time physiological indicator of stent dysfunction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Wired Pressure-Sensing Smart Biliary Stent: Bench Characterization and Feasibility in a Porcine ModelBiosensors · 2026
- Biostable wireless sensor-integrated bioresorbable stent for real-time monitoring of vascular pressure and fractional flow reserveMicrosystems & Nanoengineering · 2026
- Feasibility and Safety of a Novel Pig‐tail Inside Stent for Unresectable Malignant Hilar Biliary Obstruction: A Single‐center Case Series (With Video)DEN Open · 2026
- Functionalized bioresorbable biliary stents: materials design, coating engineering, and clinical translation.RSC advances · 2026
- The Paradigm Evolution of Biliary Stents: From Passive Drainage to Intelligent Diagnostic-Therapeutic Platforms.Acta biomaterialia · 2026
- Bioactive shape-memory biodegradable biliary stent functionalized with dHAM-Curcumin for improved biliary repair and microenvironment modulation.Biomaterials · 2026