Analytical Evaluation of Simulated Kidney-Disease Profiles: Comparison of a Portable Near-Infrared Spectrophotometer with a Benchtop FT-NIR.
Diogo Serrano, Paulo Zoio, Bernardo B Carvalho, Luís P Fonseca, Cecília R C Calado
iBB-Institute for Bioengineering and Biosciences, i4HB-The Associate Laboratory Institute for Health and Bioeconomy, IST-Instituto Superior Técnico, Universidade de Lisboa, 1049-001 Lisbon, Portugal. · Instituto de Plasmas e Fusão Nuclear, IST-Instituto Superior Técnico, Universidade de Lisboa, 1049-001 Lisboa, Portugal.
Key points
From the paper's conclusion- This study compared a portable six-channel near-infrared (NIR) analyzer (miniNIR; 1300–2100 nm) with a benchtop Fourier-transform near-infrared (FT-NIR) instrument for the discrimination and quantification of fifty-four independent concentration-defined profiles in composed fetal bovine serum solutions spanning six simulated KDIGO categories.
Abstract
Chronic kidney disease (CKD) is commonly assessed using blood biomarkers such as creatinine, urea and albumin. This study compared a portable six-channel near-infrared (NIR) analyzer (miniNIR; 1300-2100 nm) with a benchtop Fourier-transform near-infrared (FT-NIR) instrument for the discrimination and quantification of fifty-four independent concentration-defined profiles in composed fetal bovine serum solutions spanning six simulated KDIGO categories. SVM-RBF, Random Forest, XGBoost, Logistic regression and partial least-squares discriminant analysis (PLS-DA) were evaluated. For miniNIR binary 'healthy/disease' profile discrimination, SVM-RBF yielded a validation AUC of 1.00, accuracy of 0.91, sensitivity of 1.00 and specificity of 0.83. The six-class profile assignment was substantially more difficult: PLS-DA was selected by grouped nested cross-validation, but validation accuracy was 0.18, macro-F1 was 0.11 and macro-AUC was 0.77 across the evaluable classes. For miniNIR regression, validation R2 values were 0.48 for creatinine, 0.30 for urea and 0.10 for albumin. The strongest FT-NIR validation results were R2 = 0.64 for creatinine, 0.79 for urea and 0.89 for albumin, depending on the spectral window. For FT-NIR binary 'healthy/disease' profile discrimination, SVM-RBF yielded a validation AUC of 0.99, accuracy of 0.71 and a specificity of 0.47. Although the high binary AUC in this controlled concentration design for the miniNIR, it should not be interpreted as evidence of CKD-specific recognition or clinical performance. The comparison quantifies the information retained and lost when continuous FT-NIR sampling is reduced to six broad discrete channels; validation in clinically characterized human samples is required before any screening application can be considered.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Creatinine Quantification Using a Low‐Cost Portable Spectrophotometric DeviceChemistrySelect · 2026
- Classifying chronic kidney disease from non-invasive questionnaire features in adults with long-standing diabetes in Bangladesh: model development, internal validation and the limits of small-sample evidencemedRxiv · 2026
- Development of a Low-Cost Portable Exhaled Breath Ammonia Detector for Supplementary Five-Stage CKD Classification Using Embedded Threshold Logic.Sensors (Basel, Switzerland) · 2026
- Mass spectrometry-based urinary proteomic analysis in feline chronic kidney disease.Physiological reports · 2026
- Machine Learning Based Detection of Chronic Kidney Disease using Full Blood Count from a Ghanaian Hospital Population with an Integrated Web ApplicationmedRxiv · 2026
- Can Routine Clinical Variables Substitute for Urine Albumin Testing in Moderate Chronic Kidney Disease? A Machine Learning Analysis of NHANES 2015 to 2018medRxiv · 2026