NIRSLINK: A Modular Cascaded Wearable Near-Infrared Spectroscopy System for High-Speed Multi-Site Hemodynamic Monitoring.
Shuo Zhang, Kangkang Xu, Nan Zeng, Jiansong Sun, Qianrui Yang, Qianke Zeng, Zheng Ding, Yanyu Lu,
Department of Micro/Nano Electronics, Shanghai Jiao Tong University, Shanghai 200240, China. · School of Automation and Intelligent Sensing, Shanghai Jiao Tong University, Shanghai 200240, China.
Key points
From the paper's conclusion- This work develops NIRSLINK, a modular cascaded wearable NIRS system for high-speed multi-site hemodynamic acquisition.
Abstract
Wearable near-infrared spectroscopy (NIRS) enables non-invasive hemodynamic monitoring, yet conventional CW-NIRS systems suffer from limited temporal resolution, scalp-only measurement, and mandatory manual tuning to compensate for optical heterogeneity across subjects and sites. This work develops NIRSLINK, a modular cascaded wearable NIRS system for high-speed multi-site hemodynamic acquisition. Its flexible probes adopt spring-floating optics with a standardized 30 mm optode separation, integrating dual 735/850 nm LEDs, silicon photodiodes, and a two-stage closed-loop tuning algorithm. A single probe achieves a peak sampling rate of 3 kHz, and up to eight cascaded probes form 52 valid channels, with a signal-to-noise ratio (SNR) of 78.61 ± 7.03 dB and an optical dynamic range (DR) of 101.32 ± 12.41 dB. Phantom experiments verify its millisecond temporal resolution and high sensitivity to blood flow and hemoglobin variations. In vivo trials, including the Valsalva maneuver, forearm occlusion, and two-back cognitive tasks, demonstrate simultaneous recording of hemoglobin concentration shifts, pulse waveforms, and beat-to-beat pulse transit times (PTTs). NIRSLINK supports hemodynamic measurements across multiple anatomical locations, including the forehead, forearm, upper arm, and thigh, covering both cranial and peripheral body regions. The embedded auto-tuning module stabilizes signals from the forehead, forearm, and other regions within the optimal ADC range without manual adjustment, preventing signal saturation and SNR degradation. This scalable adaptive platform overcomes critical drawbacks of traditional wearable NIRS, applicable to cognitive neuroscience, non-invasive cardiovascular assessment, and ambulatory physiological monitoring, and provides design references for multi-site optical sensors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simulation-driven probe design and dual-wavelength optimization for wearable cerebral oxygenation monitoring.Biomedical optics express · 2026
- Next-Generation Wearable fNIRS: A Comprehensive Review of Bio-Instrumentation and Hardware ArchitecturesApplied Sciences · 2026
- Neckline-free open-source cap design and improved motion resistance for perioperative fNIRS procedures.Biomedical optics express · 2026
- The FetalSenseM V2: a wearable and wireless, multi-wavelength NIRS device for continuous monitoring of placental oxygenation and metabolismarXiv · 2026
- Development and validation of a low-cost, wearable near-infrared spectroscopy system for quantifying tissue water fraction in the human brain.Journal of biomedical optics · 2026
- Development of a Novel Non-Invasive Blood Pressure Measurement Technique Using Near-Infrared Spectroscopy and Deep Neural NetworksBiomedical Physics & Engineering Express · 2026