Periprosthetic Femoral Fracture in Hip Arthroplasty Using Cemented Stems: Toward a Construct-Level and Time-Dependent Perspective.
Mohamad Taufiqurrakhman, Kenzie Nathanael, Jules Deruelle, Hemant Pandit, Bernard van Duren
Leeds Institute of Rheumatic and Musculoskeletal Medicine (LIRMM), School of Medicine, University of Leeds, Leeds LS2 9JT, UK.
Key points
From the paper's conclusion- We outline a conceptual framework describing how implant migration, material response, and fracture behavior evolve with loading exposure.
Abstract
Periprosthetic femoral fracture (PFF) is an increasingly recognized complication following cemented total hip arthroplasty (THA); it is associated with substantial morbidity and mortality and often occurring several years after implantation. Although polished taper-slip (PTS) stems demonstrate favorable long-term survivorship, they have also been associated with a higher incidence of PFF, the biomechanical basis of which remains unclear. Existing studies have investigated a few key aspects of cemented THA, including implant migration, cement behavior, and fracture biomechanics. However, these factors are typically examined in isolation or under simplified conditions, and current construct-level models evaluate fracture without accounting for prior loading history. Pre-clinical testing standards relevant to fracture in cemented THA remain limited, with existing standards addressing individual components under simplified laboratory conditions rather than the integrated periprosthetic system under a physiologically relevant environment. As a result, how the implant-cement-bone construct evolves from implantation to eventual failure remains insufficiently characterized. In this study, we critically evaluate current experimental approaches and identify the absence of a time-resolved, construct-level framework as a key limitation. We outline a conceptual framework describing how implant migration, material response, and fracture behavior evolve with loading exposure. This framework may provide a foundation for both improving the mechanistic understanding of PFF and supporting the development of future pre-clinical testing strategies that can better reflect clinical loading conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Periprosthetic femoral fracture risk varies across cemented femoral stem designs: a nationwide Swiss joint registry study.International orthopaedics · 2026
- Early Nontraumatic Periprosthetic Femoral Fracture After Total Hip Arthroplasty: A Case Report.Case reports in orthopedics · 2026
- Maximizing stem body size may reduce the risk of postoperative periprosthetic fractures : an analysis of 2,757 CPT femoral components.Bone & joint open · 2026
- Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026
- FP12.5 Long-Term Periprosthetic Fracture Risk After Hip Resurfacing Arthroplasty: A 25-Year Analysis Supporting Femoral Neck PreservationJournal of Hip Preservation Surgery · 2026
- Cemented total hip arthroplasty reduces early complications: a Japanese nationwide propensity-matched studyArchives of Orthopaedic and Trauma Surgery · 2026