The Impact of the Impella RP® Device on a Failing Right Heart: A Modelling and Simulation Approach to Ascertain Its Potential.
Beatrice De Lazzari, Angela Richter, Christoph Nix, Roberto Badagliacca, Annalisa Pitino, Mercedes Gori, Gianmarco Scoccia, Massimo Capoccia,
Department of Movement, Human and Health Sciences, University of Rome "Foro Italico", 00135 Rome, Italy. · J&J MedTech Heart Recovery, Neuenhofer Weg, 52074 Aachen, Germany.
Key points
From the paper's conclusion- The Impella RP® device has the potential to become a valid option for RV support based on current experimental and simulation data. Although already in use, further evaluation in the clinical setting will likely confirm its potential and lead to a more routinely application for RV support.
Abstract
BACKGROUND: Indications for right ventricular assist device (RVAD) insertion include right heart failure after implantation of a left ventricular assist device or early graft failure following heart transplantation. This study aimed to investigate how the upstream and downstream circulatory network interacts with the Impella RP® device. METHODS: A numerical model of the Impella RP® device was implemented within the CARDIOSIM© software platform for this study. In the numerical configuration, the RVAD aspirated blood from either the right atrium (RA-PA connection) or the right ventricle (RV-PA connection) and delivered it to the pulmonary artery. Only RA-PA connection is the currently used setting for the Impella RP® device in clinical practice. Based on right ventricular (RV) decompression and total flow, our study may help define the need for a direct RV-unloading Impella RP®. RESULTS: The simulations showed that while activating the RVAD in RA-PA mode, regardless of its rotational speed, the mean pulmonary artery pressure (PAP) percentage change was higher than that in the unsupported condition when the mean systemic venous pressure (SVP) and the pulmonary artery wedge pressure (PAWP) were both set to 20 [mmHg]. When RV-PA connection was applied, a similar trend was observed, although the PAP percentage changes were approximately halved compared to those when the RA-PA connection was applied. CONCLUSIONS: The Impella RP® device has the potential to become a valid option for RV support based on current experimental and simulation data. Although already in use, further evaluation in the clinical setting will likely confirm its potential and lead to a more routinely application for RV support.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biventricular Pressure-Volume Loop Analysis in an LVAD Patient Optimized With Impella RP Flex.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Hemodynamic Effects of Three Different Mechanical Circulatory Support Strategies in Isolated Acute Right Heart Failure.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Right ventricular assist device support during left ventricular assist device implantation: indications, timing, device selection, and perioperative managementFrontiers in Cardiovascular Medicine · 2026
- Biventricular Pressure-Volume Loop Assessment of Impella 5.5 in an Acute ST-Segment-Elevation Myocardial Infarction Shock Model Unmasks Diminished Right Ventricular Reserve.Journal of the American Heart Association · 2026
- Early Right Heart Failure After Left Ventricular Assist Device Implantation: A Narrative Review of Definitions, Risk Prediction, and Management.Cardiovascular therapeutics · 2026
- Acute Right Ventricular Failure in Animal Models: Evidence, Limitations, and Future Directions.Artificial organs · 2026