A Low-Cost Fourier-Based Dual-Pump Mock Circulatory Loop for Arterial Pressure Waveform Reproduction.
Victor K Tsui, Kouhyar Tavakolian, Daniel Ewert
Department of Biomedical Engineering, University of North Dakota, Grand Forks, ND 58202, USA. · Department of Electrical and Computer Engineering, Old Dominion University, Norfolk, VA 23529, USA.
Key points
From the paper's conclusion- These findings demonstrate that the proposed Fourier-based hMCL provides an accurate, reproducible, and cost-effective platform for arterial pressure waveform reproduction and early-stage cardiovascular device testing.
Abstract
Cardiovascular diseases are among the leading causes of death worldwide, creating a need for reliable hemodynamic testing platforms for cardiovascular device evaluation and therapeutic development. Mock circulatory loops (MCLs) provide effective in vitro platforms for reproducing physiological and pathological pressure conditions; however, many existing systems remain limited in programmability, waveform fidelity, standardization, and affordability. This study presents a cost-effective dual-pump hybrid MCL (hMCL) that independently regulates baseline and pulsatile pressure components using Fourier-based harmonic synthesis. Reference arterial pressure waveforms were decomposed into harmonic components, and their amplitudes, frequencies, and phase shifts were converted into pulse-width-modulation signals for dual-pump actuation. Validation using five physiological and virtual arterial pressure waveform datasets demonstrated MAP errors of ≤1.3%, RMSE values of 9.1-24.1 mmHg, and feature errors ≤ 2.8% for systolic, diastolic, and dicrotic notch pressures. Distribution analysis showed small median pressure differences and low variability across waveform characteristics, while repeated hMCL operation demonstrated pressure variations below 3 mmHg. These findings demonstrate that the proposed Fourier-based hMCL provides an accurate, reproducible, and cost-effective platform for arterial pressure waveform reproduction and early-stage cardiovascular device testing. The present validation is pressure-focused; flow waveform fidelity and pressure-flow interactions were not independently characterized.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Non-invasive estimation of brachial pressure waveforms via oscillometric transfer function analysis using cardiovascular hardware simulatorScientific Reports · 2026
- Hemadyne: accordion-inspired perfusion for microphysiological systemsNature Communications · 2026
- Soft-closure bileaflet mechanical heart valve design and hydrodynamic evaluation: an integrated iValve framework.Journal of biomechanics · 2026
- An ECG-PPG Physiological Signal Emulator for Calibration and Validation of Cardiovascular Monitoring Devices.Sensors (Basel, Switzerland) · 2026
- Evolutionary Optimization of Extended Kalman Filter Hyperparameters for Adaptive Data-Driven ABP Prediction Framework Inspired by Cardiovascular Digital-Twin PrinciplesAlgorithms · 2026
- Open-loop system identification of an AC motor-driven arterial flow generation platform: vessel-specific models and a parsimonious common ARX model.Medical engineering & physics · 2026