Automatic Detection of Extracochlear Electrodes in Cochlear Implants Using Electric Field Imaging: Bench and Clinical Validation.
Mehrangiz Ashiri, Tony Spahr, Chen Chen, Azret Botash, Ashish Mehta, Patrick Boyle, Manohar Bance, Daniele Borsetto,
Research and Technology, Advanced Bionics LLC, Valencia, CA 91355, USA. · Department of Clinical Neurosciences, University of Cambridge, Cambridge CB2 0QQ, UK.
Key points
From the paper's conclusion- This EFI-based algorithm may provide a useful screening tool for EEs using routine clinical measurements, supporting intra-operative and post-operative detection while reducing reliance on imaging and thereby aiding clinical decision-making. The method can be integrated into existing clinical fitting software workflow and complements recent work on EFI-based tip fold-over detection. The observed sensitivity and specificity suggest that this approach may provide a valuable tool for identifying EEs in situations where imaging is unavailable or impractical.
Abstract
OBJECTIVES: Extracochlear electrodes (EEs), defined as electrode contacts located outside the cochlea, can degrade cochlear implant (CI) performance. We present and validate an Electric Field Imaging (EFI)-based algorithm that helps detect EEs and can be used intra- or post-operatively. METHODS: An algorithm for detecting EEs from EFI data was developed. Validation and testing were performed using three datasets: (a) bench models with known EE conditions (saline and resistive load model), (b) clinical EFI recordings from CI recipients with imaging confirmation (CT or plain X-ray), and (c) EFI recordings without imaging. Primary outcomes included the ability to differentiate extracochlear conditions from fully inserted electrode arrays, as well as the concordance between the number of EE contacts identified by the algorithm and those determined from controlled bench configurations or available imaging data. RESULTS: The algorithm reliably differentiated full insertion from EE conditions on bench models and clinical EFI recordings from CI recipients with imaging confirmation. In the imaging-confirmed clinical cohort (6 EE cases among 226 CI recipients), the algorithm achieved 100% sensitivity and specificity. In clinical cases without imaging, the algorithm flagged 3.94% as having extracochlear electrodes, within the range of prevalence reported in the literature. Furthermore, the lateral-wall electrode arrays were associated with a significantly higher extracochlear electrode occurrence compared to the pre-curved arrays. CONCLUSIONS: This EFI-based algorithm may provide a useful screening tool for EEs using routine clinical measurements, supporting intra-operative and post-operative detection while reducing reliance on imaging and thereby aiding clinical decision-making. The method can be integrated into existing clinical fitting software workflow and complements recent work on EFI-based tip fold-over detection. The observed sensitivity and specificity suggest that this approach may provide a valuable tool for identifying EEs in situations where imaging is unavailable or impractical.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Automated, Radiation-Free Detection of Cochlear Implant Electrode Array Migration Using Routine Impedance Telemetry.Journal of medical systems · 2026
- Impedance-based tracking of a cochlear implant electrode reliably detects successful array insertion.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Correlating Electrically Evoked Compound Action Potential Thresholds with Cochlear Implant Position Using Validated Computed Tomography-Based Electrode Localization.The journal of international advanced otology · 2026
- Acoustic Masking of Electric Stimulation at Basal and Extra-Cochlear SitesbioRxiv · 2026
- Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Electrically evoked compound action potentials in adult and pediatric cochlear implant users: I. Cochlear nerve responsiveness and electrode-neuron interface index.Hearing research · 2026