European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery

Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation.

Annett Franke-Trieger, Thomas Zahnert, Katrin Murrmann, Susen Lailach

Saxonian Cochlear Implant Centre, Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine (and University Hospital), Technische Universität Dresden, Fetscherstraße 74, 01307, Dresden, Germany. annett.trieger@ukdd.de. · Saxonian Cochlear Implant Centre, Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine (and University Hospital), Technische Universität Dresden, Fetscherstraße 74, 01307, Dresden, Germany.

Observational

Key points

From the paper's conclusion
  • This variability, captured by the LID E12 measurement, demonstrates that the real-life result of the electrode placement often differs from the assumed placing, and that this factor should be considered when planning electrode positioning.

Abstract

OBJECTIVE: Accurate prediction of the angular insertion depth (AID) is essential for optimising cochlear implant (CI) outcomes. However, discrepancies between preoperative predictions and postoperative results persist. In this study, we evaluated systematically the accuracy of equation-based AID prediction to identify the main contributors to prediction errors. METHODS: In this retrospective study, preoperative AID predictions for various MED-EL electrode arrays were generated by using an equation-based method, assuming full insertion with the electrode stopper at the round window. These predictions were compared with postoperatively measured AIDs obtained by digital volume tomography. The primary variable investigated was the linear insertion depth of the electrode closest to the basal position (LIDE12), defined as its distance from the round window. RESULTS: The LIDE12 ranged from - 13.8 mm to 11.0 mm; negative values indicate incomplete insertions, which were observed in 9% of cases. Between predicted and measured AID a median difference of - 11° (interquartile range 74°) was observed. After correcting the prediction by incorporating the measured LIDE12 instead of assuming insertion to the stopper, the median difference shifted to + 35° with a markedly reduced interquartile range (37°). CONCLUSION: While current AID prediction models generally show clinically acceptable accuracy, considerable discrepancies can still occur. One major contributing factor is the actual surgical placement of the electrode array. This variability, captured by the LIDE12 measurement, demonstrates that the real-life result of the electrode placement often differs from the assumed placing, and that this factor should be considered when planning electrode positioning.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation — Science Skim