Design and Exploratory Characterization of a 3D-Printed PETG Laryngoscope for Simulation-Based Medical Training
Darwin Castillo, Vasudevan Lakshminarayanan, Leonel Vasquez-Cevallos
Key points
From the paper's conclusion- Reproduced force averaged 12.58 ± 4.18 kgf (123.38 ± 40.97 N), ranging from 5.0 to 20.0 kgf.
Abstract
This study characterized the design, assembled mass, and analytical bending behavior of a non-commercial glycol-modified polyethylene terephthalate (PETG) laryngoscope for supervised academic simulation. A secondary cross-sectional analysis described responses from 314 novice medical students. After manikin laryngoscopy, each participant compressed a separate dynamometer with the same hand to reproduce the greatest perceived manual effort during the maneuver. This outcome is an instruction-defined recalled-effort response, not a measurement of laryngoscope loading or educational effectiveness. The prototype had a single assembled-mass reading of 95 g; two photographed commercial configurations had readings of 454 and 522 g, without established functional equivalence. Analytical cross-section integration yielded a minimum load-to-stress ratio of 2.483 N/MPa under homogeneous, fully bonded elastic bending assumptions. This ratio characterizes modeled stress demand; it does not establish structural capacity. Reproduced force averaged 12.58 ± 4.18 kgf (123.38 ± 40.97 N), ranging from 5.0 to 20.0 kgf. These observations document the prototype configuration and the distribution of responses to the reproduction task. Missing manufacturing settings limit reproducibility, while mechanical capacity, actual maneuver forces, functional performance, and educational effectiveness require separate validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Flexible Neuroendoscopic Training System Using a 3-dimensional Printed Brain Model with Side-view Visualization of Ventricular Wall Contact.Neurologia medico-chirurgica · 2026
- Development and Evaluation of a 3D Printed Knee Model for Simulating Partial Meniscus ResectionCurrent Directions in Biomedical Engineering · 2026
- Comparative Ergonomic Assessment of Microscopic, Endoscopic, and RobOtol®-Assisted Ear Surgery: A Prospective Pilot Study.Journal of clinical medicine · 2026
- Development and evaluation of the first fully 3D-printed 5-mm laparoscopic grasper for surgical training.Surgical endoscopy · 2026
- Distance Hands‐On Training Versus Video Instruction for Flexible Laryngeal Mask Airway Management: A Protocol for a Multicenter Randomized Controlled TrialActa Anaesthesiologica Scandinavica · 2026
- Design and implementation of a paradigm-specific cervical manipulation simulator with objective force-angle feedback.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026