Effect of a robotic insole-type active assist device on horizontal ground reaction force and center-of-pressure stability during stepping in patients with medial knee osteoarthritis.
Taku Itami, Takaaki Aoki, Haruhiko Akiyama
Department of Rehabilitation and Orthopedics, Gifu University, Gifu, Japan.
Key points
From the paper's conclusion- These findings suggest that dynamic ankle alignment correction using a robotic insole-type device may reduce horizontal loading variability and center-of-pressure sway during stepping without substantially altering vertical loading.
Abstract
An insole-type active assist device has been developed as a robotic system to dynamically correct ankle alignment at heel contact in patients with medial knee osteoarthritis. Although our previous feasibility study demonstrated that the device could be safely used during an on-the-spot stepping task, its effects on loading behavior remain unclear. This study aimed to investigate whether dynamic ankle alignment correction using the device alters horizontal ground reaction force variability and center-of-pressure sway during stepping. Force-plate data obtained from six ambulatory patients with medial knee osteoarthritis were analyzed as a secondary biomechanical analysis. Each participant performed repeated stepping trials under two conditions: a non-control condition, in which the device was worn without motor control, and a control condition, in which heel-eversion assistance was provided. Stance phases were extracted using vertical ground reaction force, and horizontal ground reaction force components, center-of-pressure sway measures, peak vertical ground reaction force, and vertical impulse were calculated. Compared with the non-control condition, the control condition reduced the ranges of anterior-posterior and medial-lateral ground reaction force components in all participants. Center-of-pressure rectangular sway area also decreased consistently, and trajectory length tended to decrease, whereas peak vertical ground reaction force and vertical impulse showed little change. These findings suggest that dynamic ankle alignment correction using a robotic insole-type device may reduce horizontal loading variability and center-of-pressure sway during stepping without substantially altering vertical loading. Because this study was based on a small sample and an exploratory secondary analysis, the findings should be interpreted as preliminary biomechanical evidence rather than evidence of clinical effectiveness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ankle Foot Orthosis Intervention Improves the Ground Reaction Forces During Walking in Patients with Peripheral Artery Disease (Randomized Clinical Trial).Actuators · 2026
- Directional contrasts in center of pressure and external knee adduction moment during stepping with wedge insoles, verbal instructions, and perceptual plantar cues: An exploratory within-subject study.Gait & posture · 2026
- Adaptive Gait Biofeedback With Participant-Held-Out Modeling and Participant-Specific Updating in Chronic Ankle InstabilityarXiv · 2026
- An Adaptive Knee Brace Is Associated With Improved Dynamic Stabilization Compared With a Standard Passive Brace in Anterior Cruciate Ligament-Deficient Men During a Functional Hop Task.Arthroscopy, sports medicine, and rehabilitation · 2026
- Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre-Post Motion Analysis Study.Bioengineering (Basel, Switzerland) · 2026
- In vivo quantification of knee-orthosis valgus-corrective moment during gait: Effects of medial strut length on external knee adduction moment in medial knee osteoarthritis.PloS one · 2026