Performance evaluation of vacuum blood collection tubes in multicenter clinical laboratories: a real-world study from Southwest China.
Changrui Sun, Chunbao Xie, Xuelong Xie, Li Li, Jian Xu, Jing Ren, Dan Zheng, Juan Du,
Department of Laboratory Medicine, Sichuan Academy of Medical Sciences and Sichuan Provincial People's Hospital, Chengdu, China. · Department of Clinical Laboratory, The Second People's Hospital of Yibin·West China Yibin Hospital, Sichuan University, Yibin, China.
Key points
From the paper's conclusion- Southwest China's routinely used vacuum blood collection tubes exhibit critical performance gaps, particularly in fill-volume accuracy and stopper integrity. Routine verification of high-risk indicators, site-adaptive batch-specific verification for high-altitude facilities, and manufacturing improvements (e.g., stopper material optimization) are recommended. This study provides quality benchmarks for regional test result mutual recognition and pre-analytical quality improvement.
Abstract
BACKGROUND: Pre-analytical errors account for 60-70% of total laboratory errors, with vacuum blood collection tube quality being a key contributor. Southwest China's complex terrain (plain/plateau areas) and demand for inter-hospital test result mutual recognition highlight the need for region-specific performance data. METHODS: This multicenter real-world study evaluated 10 blood collection tube brands across 10 hospitals (plain/plateau areas) in Sichuan-Chongqing. Nine physical performance indicators (appearance, tube strength, fill volume, stopper shedding, hemolysis, fibrin adhesion, separation gel integrity, anticoagulation efficacy, and tube-body integrity) were verified against BD tubes, following CLSI GP34-A2 and WS/T 224-2018, with additional clinical consistency analyses for hematological, coagulation, and biochemical parameters. Statistical analysis included linear regression (R 2) and relative/comparison bias. RESULTS: Widespread performance deficiencies were observed. At the hospital level, fill-volume non-compliance affected 78.3% (7/9) of plain-area hospitals, while at the tube-batch level, 20.0% (1/5) of batches failed at the single plateau site. For stopper shedding, 83.3% (25/30) of tube batches across all sites exceeded the acceptable threshold, corresponding to 80% (8/10) of hospitals having at least one non-compliant tube type. Tube appearance, structural strength, and separation gel integrity were universally compliant across all sites. CONCLUSION: Southwest China's routinely used vacuum blood collection tubes exhibit critical performance gaps, particularly in fill-volume accuracy and stopper integrity. Routine verification of high-risk indicators, site-adaptive batch-specific verification for high-altitude facilities, and manufacturing improvements (e.g., stopper material optimization) are recommended. This study provides quality benchmarks for regional test result mutual recognition and pre-analytical quality improvement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishing denominator-consistent Six Sigma benchmarks for venous blood specimen rejection: a multi-center retrospective quality-indicator study using 21 million specimens in Western China.Frontiers in medicine · 2026
- Evaluation of Pre-Analytical Performance Using IFCC Quality Indicators, Six Sigma Metrics and Root Cause Analysis in a Biochemistry Laboratory-A Retrospective Study.Diagnostics (Basel, Switzerland) · 2026
- Partitioning All-Cause Pre-Analytical Laboratory Rejections: A 25-Month Retrospective Audit of Specimen Quality Failures with Six Sigma, IFCC, and Pareto Analysis.Journal of clinical medicine · 2026
- Evaluation of UNI-VAC serum separator, ethylenediaminetetraacetic acid, sodium citrate, and sodium fluoride tubes: A comparison with Becton, Dickinson and Company Vacutainer tubes.PloS one · 2026
- B-046 Automated De-Capping Compatibility of V-PRO® Blood Collection Tubes in a Total Laboratory Automation SystemClinical Chemistry · 2026
- Spectrum of Preanalytical Laboratory Errors in Clinical Biochemistry Laboratory in a Tertiary Care Center in Nepal.Kathmandu University medical journal (KUMJ) · 2025