Comparison of a novel portable wireless multichannel EEG with inbuilt seizure detection algorithm to a conventional EEG in a term neonatal cohort
Anne L. Murray, Sergio Morant Gálvez, Andreea M. Pavel, Aisling Fanning, Vicki Livingstone, Alison O’Shea, Mark E. O’Sullivan, Brian H. Walsh
Key points
From the paper's conclusion- The secondary outcome was to describe the performance of the Neonatal Seizure Detection Algorithm (NSDA) compared to seizure annotation by an expert reviewer.
Abstract
Background Electroencephalography (EEG) has become increasingly important in neonatal neurocritical care. The use of continuous multichannel EEG (cEEG) can be limited in some settings by lack of expertise, personnel and resources. Alternatives to traditional cEEG devices need to be explored to allow more infants to avail of appropriate monitoring when indicated. The aim of this study was to assess the correlation of the signal recorded with a portable wireless EEG device with that of a conventional EEG monitor. Methods This was a single-centre, prospective method comparison study (NCT07508670). Near-term or term neonates requiring multichannel EEG were eligible for inclusion. EEG was simultaneously recorded on conventional (12 channel) and investigational (8-channel) devices using a single set of split electrodes. The primary outcome was EEG signal similarity between novel and conventional devices, assessed using Pearson correlation coefficients (PCCs). The secondary outcome was to describe the performance of the Neonatal Seizure Detection Algorithm (NSDA) compared to seizure annotation by an expert reviewer. Results 17 neonates, median gestation (interquartile range [IQR]) 39 + 2 weeks (38 + 4-40 + 6), were enrolled. A total of 342 h of overlapping EEG were analysed, median overlap in hours per recording was 19.48 (IQR: 14.91–21.70). The primary outcome analysis demonstrated that EEG signals were highly correlated between devices, median PCC=0.966 (95% confidence interval [CI]: 0.920 to 0.984). For the secondary outcome, while only one infant had seizures while connected to the investigational device, the NSDA identified all recorded seizures ( n = 4), with 0.013 (95% CI 0.004–0.025) false detections per hour (FD/h). Conclusion The investigational device recorded EEG comparable to a conventional monitor. Its integrated NSDA identified all seizures in a single infant with infrequent false detections, but larger studies are required to test its clinical performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- U-NeXt: General-purpose event detection for neonatal EEGarXiv · 2026
- Automated EEG seizure detection in neonates with hypoxic–ischemic encephalopathy using temporal and morphological modelingFrontiers in Pediatrics · 2026
- Developmental domain shifts in epileptiform discharges: multidomain EEG differences between neonates and adultsFrontiers in Neuroscience · 2026
- Amplitude-integrated electroencephalogram (aEEG): A bedside window into newborn brain function in the NICU.Seminars in fetal & neonatal medicine · 2026
- Expert-level detection of low-voltage intervals in EEG as a robust biomarker of neonatal encephalopathyarXiv · 2026
- Clinical Implementation of Amplitude-Integrated Electroencephalography for Neonatal Seizure Detection and Prognostication: First National Experience From North Macedonia.Pediatric neurology · 2026