Modelling image quality over extended [18F]FDG uptake times for low-dose positron emission mammography.
Elliott De Benetti, Tobias Preckel, Apichart Linhananta, Brandon Baldassi, Borys Komarov, Harutyun Poladyan, Vivianne Freitas, Alla Reznik,
Department of Physics, Lakehead University, Thunder Bay, ON, Canada. · Department of Medical Engineering, Pforzheim University, Pforzheim, Germany.
Key points
From the paper's conclusion- Extending uptake times beyond the 1-hour clinical standard significantly improved CNR across all doses, with comparable proportional gains. Relative to 1 hour post-injection, CNR increased by approximately 30% at 1.5 hours, 53% at 2 hours, and 77% at 3 hours, although uncertainty widened at 3 hours and beyond, particularly at the lowest dose (1 mCi). Relatively small variations in uptake time were also found to produce substantial changes in the standardized uptake value (SUV).
Abstract
INTRODUCTION: This study investigated the effect of [18F]fluorodeoxyglucose ([18F]FDG) uptake time on image quality in low-dose positron emission mammography (PEM) using kinetic modelling. METHODS: Fifteen malignant lesions from thirteen women newly diagnosed with breast cancer were imaged using low-dose [18F]FDG PEM. Tumour and normal breast tissue time-activity curves (TACs) were generated using a population-based arterial input function (AIF) and a two-tissue compartmental model. Modelling was performed at [18F]FDG doses of 1, 2, 5, and 10 mCi (37, 74, 185, and 370 MBq), and image quality was assessed using contrast-to-noise ratio (CNR) over a 6-hour uptake period. RESULTS: Extending uptake times beyond the 1-hour clinical standard significantly improved CNR across all doses, with comparable proportional gains. Relative to 1 hour post-injection, CNR increased by approximately 30% at 1.5 hours, 53% at 2 hours, and 77% at 3 hours, although uncertainty widened at 3 hours and beyond, particularly at the lowest dose (1 mCi). Relatively small variations in uptake time were also found to produce substantial changes in the standardized uptake value (SUV). DISCUSSION: These findings demonstrate that extending uptake times by even 30 minutes beyond the 1-hour clinical standard may significantly enhance image CNR for low-dose [18F]FDG PEM. The sensitivity of SUV to small variations in uptake time supports the need to closely match uptake time between baseline and follow-up scans in treatment response assessment. Overall, these modelling results encourage further clinical research into optimizing uptake time protocols, which may help support the broader adoption of low-dose PEM in future work. CLINICAL TRIALS REGISTRATION: https://clinicaltrials.gov/study/NCT03520218, identifier NCT03520218.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimization of PET acquisition time for thoracic lesions in 18F-FDG PET/MRI: a retrospective analysis.European radiology experimental · 2026
- Technical evaluation of delayed-phase SUV estimation using four-phase whole-body dynamic 18F-FDG PET/CT: a phantom and clinical study.Radiological physics and technology · 2026
- Deep progressive reconstruction for low-dose ¹⁸F-FDG PET/CT imaging in pediatric patients.Annals of nuclear medicine · 2026
- Test-Retest Repeatability of [18F]FDG Uptake by Solid Tumors on PET/MRI: A Prospective, Multicenter Study Using the QIBA Protocol.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026
- Optimization of patient dose and image quality for positron emission tomography-computed tomography.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Serial changes in myocardial fluorodeoxyglucose uptake on routine oncologic PET during breast cancer chemotherapy: are they clinically relevant?Nuclear medicine communications · 2026