Optimization of nested U-Net-based noise reduction for low-field brain MRI: effects of pruning level and patient posture
Hajin Kim, Chang-Soo Yun, Changheun Oh, Jun-Young Chung, Youngjin Lee, Gun Choi
Key points
From the paper's conclusion- This study evaluated the feasibility of improving the image quality of 0.25-T low-field brain MRI using a nested U-Net-based noise reduction model and optimized the model through pruning methods.
Abstract
Introduction Magnetic resonance imaging (MRI) is widely used for the diagnosis and evaluation of brain disorders as it enables superior soft-tissue contrast and the detection of fine anatomical structures and pathological lesions. Recently, low-field MRI has attracted increasing attention owing to its low cost, improved accessibility, portability, and open-bore design; however, its inherently low signal to noise ratio limits image quality and clinical utility. This study evaluated the feasibility of improving the image quality of 0.25-T low-field brain MRI using a nested U-Net-based noise reduction model and optimized the model through pruning methods. Methods Four pruning levels, ranging from L1 (the shallowest and most pruned configuration with the fewest parameters) to L4 (the full-depth nested U-Net without pruning), were assessed using the peak signal to noise ratio (PSNR), structural similarity index measure (SSIM), and mean absolute error (MAE). Computational efficiency was evaluated using training time, inference time, and the number of parameters. Results The pruning level significantly affected all metrics, with L3 achieving the highest PSNR and the lowest MAE and an SSIM comparable to the other levels. Compared with L4, L3 reduced the number of parameters by 75.6%, training time by 30.1%, and inference time by 24.3%, while improving PSNR by 1.75% (an increase from 37.11 to 37.76 dB) and MAE by 2.42% (a decrease), with a comparable SSIM (0.908 vs. 0.914). L3 also showed the smallest posture-related performance difference and outperformed conventional filters (Gaussian, median, Wiener, and non-local means) and a standard U-Net in PSNR and MAE with 28% of the parameters of the U-Net. Discussion Overall, these results suggest that an intermediate pruning level (L3) provides a favorable trade-off between noise reduction performance and computational efficiency for low-field brain MRI, warranting further validation in clinical populations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Deep learning-based image domain reconstruction of undersampled low-field brain MRI: A comparative benchmark study of U-Net architectures.Magnetic resonance imaging · 2026
- A Dual-Stream Regulated Reconstruction and Segmentation Network with Hierarchical Artifact-Prior Modeling for Ultra-Low-Field Pediatric NeuroimagingarXiv · 2026
- Noise removal and intensity inhomogeneity correction in brain MRI: A comprehensive review from classical filters to transformer and diffusion models.Magnetic resonance imaging · 2026
- On the accuracy of nonlinear image registration in portable low-field 3D brain MRI.Scientific reports · 2026
- A Guarded Hybrid Pipeline for Rigid FLAIR–T1 MRI Registration: Learned Initialisation of Iterative Mutual-Information RefinementBioMedInformatics · 2026