Diagnostic accuracy of urine test strips for CSF analysis: high sensitivity in detecting blood, low sensitivity but high specificity in detecting inflammation.
Lina Marie Albers, Franz Felix Konen, Erda Atakan, Konstantin Fritz Jendretzky, Sandra Nay, Lea Grote-Levi, Kurt-Wolfram Sühs, Thomas Skripuletz,
Hannover Medical School, Department for Neurology, Hannover, Germany.
Key points
From the paper's conclusion- Therefore, while UTS may support early bedside decision-making, they cannot replace standard laboratory diagnostics.
Abstract
BACKGROUND: Rapid cerebrospinal fluid (CSF) analysis is crucial for diagnosing and treating time-critical neurological diseases. However, thorough laboratory diagnostics are time-intensive or often unavailable in resource-limited settings. Urine test strips (UTS) may provide a simple bedside alternative, but evidence remains limited and largely restricted to single manufacturers. This study evaluated the diagnostic performance of UTS for CSF analysis and compared different manufacturers. METHODS: In this prospective study, 285 CSF samples from patients with different neurological diseases were analyzed using UTS from five manufacturers. Strip results for leukocytes, erythrocytes, protein, and glucose were compared across manufacturers and against laboratory measurements. Diagnostic accuracy for detecting blood in CSF, pleocytosis, and for differentiating infectious from non-infectious conditions was assessed. RESULTS: For erythrocyte detection, the UTS showed high sensitivity (86-95%) and negative predictive value (NPV, 92-97%). In contrast, for CSF pleocytosis, specificity and positive predictive value (PPV) were high (100%), whereas sensitivity was low (11%). UTS underestimated low leukocyte counts and showed variable performance for protein detection. For differentiating infectious from non-infectious conditions, leukocytes and glucose demonstrated high specificity and PPV. Combining parameters (≥ 2/3 positive: leukocytes, protein, glucose) yielded consistently high specificity (98%) and PPV (80-85%), but low sensitivity (13-15%) and NPV (63%). No relevant differences between manufacturers were observed. CONCLUSION: UTS provide a rapid, low-cost bedside tool for initial CSF assessment, with particular strength in reliably excluding blood contamination. In contrast, their utility for detecting inflammatory or infectious conditions is limited due to low sensitivity. Therefore, while UTS may support early bedside decision-making, they cannot replace standard laboratory diagnostics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic Accuracy of Urinary Reagent Strip Applied to Cerebrospinal Fluid Analysis for the Detection of Bacterial Meningitis in Children.Mymensingh medical journal : MMJ · 2026
- B-001 Limitations of Urine Dipstick Testing and the Clinical Value of Creatinine-Normalized Quantitative Urinalysis in Urologic and Renal Disease DetectionClinical Chemistry · 2026
- A-076 Evaluation of the Sysmex UF-5000 WBC parameter to detect urethral inflammation in remnant male urine specimens: Implications for ruling-in urethritisClinical Chemistry · 2026
- A-110 The Jiddu Platform: Automated One-Hour Testing for Bacterial Detection and Antibiotic Sensitivity in Clinical SamplesClinical Chemistry · 2026
- Evaluation of the Sysmex UF-5000 automated urine flow cytometer for detection of urethral inflammation and clinical urethritis in men.Journal of clinical microbiology · 2026
- Diagnostic accuracy of cerebrospinal fluid gram stain for bacterial meningitis in patients with a suspected central nervous system infection.Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases · 2026