Digital Design of Kurtosis-Controlled Ti-6Al-4V Lattices for Patient-Specific Orthopedic Implants: A Computational Framework.

Marzhan Sadenova, Boris Syrnev, Bagdat Azamatov

Center of Excellence «VERITAS», D. Serikbayev East Kazakhstan Technical University, Ust-Kamenogorsk 070000, Kazakhstan.

Computational

Key points

From the paper's conclusion
  • A GAN-CAD-FEA workflow reproduced the analytical moduli to within 7% across six design cases.

Abstract

Porous Ti-6Al-4V lattice implants combine high specific strength, osseointegrative porosity, and compatibility with additive manufacturing, but conventional stiffness tuning through porosity, pore size, or unit-cell topology compromises biological pore requirements. This study presents a computational design framework in which structural kurtosis, the normalized interlayer offset between neighboring layers of a periodic cubic lattice, regulates elastic response at fixed global porosity. Closed-form expressions for the effective modulus are derived from first principles: the aligned configuration from the axial load-bearing area fraction, and the interlayer-shifted configuration from Euler-Bernoulli beam theory for guided-end connecting members. The derivations reproduce the Gibson-Ashby exponents n = 1 and n = 2, replacing the previously asserted power law, and a calibrated one-parameter interpolation bridges intermediate offsets. At 65% porosity, the effective modulus falls from 16.5 GPa in the aligned lattice to 2.64 GPa in the shifted lattice. A local-yield analysis based on peak bending curvature gives recoverable elastic strains of 1.37% at 89% porosity and 0.68% at 65%; the compliance-based values of 20.5% and 5.12% are kinematic upper bounds that neglect plastic hinging. A prefactor-free benchmark shows that obtaining the same 6.25-fold reduction by increased porosity alone would require 85.9-94.4% porosity and 0.17-0.28 mm struts, outside the osseointegration window and the resolution of selective laser melting. A GAN-CAD-FEA workflow reproduced the analytical moduli to within 7% across six design cases. All results are analytical and numerical; no specimens were fabricated or tested, and experimental validation remains required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Digital Design of Kurtosis-Controlled Ti-6Al-4V Lattices for Patient-Specific Orthopedic Implants: A Computational Framework — Science Skim