Real-world validation of an AI-based non-mydriatic fundus camera in the diagnosis of diabetic retinopathy in Liberia.
Vickie Fayiah Sam, Mohammad Rizwan, John Chiemerie Douglas, Amrit Jha, Sirisha Senthil, Padmaja Kumari Rani
LVPEI-Liberia Eye Center, John F. Kennedy Memorial Hospital, Monrovia, Liberia. · Kalathur Venugopal Reddy Glaucoma Institute, L V Prasad Eye Institute, Hyderabad, India.
Key points
From the paper's conclusion- In this prospective real-world study, the offline AI-based non-mydriatic fundus camera showed high sensitivity and specificity for detecting referable diabetic retinopathy among participants with at least one gradable eye in a general outpatient setting in Liberia. These findings support its feasibility for AI-assisted detection of referable DR in resource-limited settings.
Abstract
INTRODUCTION: Diabetic retinopathy (DR) is a leading cause of preventable vision loss, yet access to retinal screening remains limited in low-resource settings such as Liberia. This study evaluated the real-world diagnostic performance of an offline artificial intelligence (AI)-enabled smartphone-based non-mydriatic fundus camera for detecting referable diabetic retinopathy in a routine general outpatient screening setting in Liberia. METHODS: This prospective study was conducted at the General Outpatient Department of John F. Kennedy Medical Center, Monrovia, Liberia. Adults aged 18 years and older were recruited through convenience sampling and underwent non-mydriatic fundus imaging using the Remidio Fundus on Phone (FOP-NM-10) integrated with an offline AI system. Fundus images were independently assessed by two trained optometrists using the International Clinical Diabetic Retinopathy (ICDR) classification, with disagreements adjudicated by a senior retinal ophthalmologist. Referable DR was defined as moderate non-proliferative DR, severe non-proliferative DR, or proliferative DR. Diagnostic performance was evaluated at the participant level among participants with at least one gradable eye. RESULTS: A total of 1,612 participants were enrolled, of whom 1,357 were included in the final analytical cohort. Of these, 1,198 (88.3%) had at least one gradable eye and were included in the diagnostic-performance analysis. Eleven participants were classified as having referable DR by the human reference standard. The AI system correctly classified 10 of these 11 participants, yielding a sensitivity of 90.9% (95% CI, 58.7%-100%). Among 1,187 participants without referable DR, 1,183 were correctly classified as negative, resulting in a specificity of 99.7% (95% CI, 99.1%-99.9%). The positive predictive value was 71.4% (95% CI, 41.9%-91.6%), and the negative predictive value was 99.9% (95% CI, 99.5%-99.9%). CONCLUSIONS: In this prospective real-world study, the offline AI-based non-mydriatic fundus camera showed high sensitivity and specificity for detecting referable diabetic retinopathy among participants with at least one gradable eye in a general outpatient setting in Liberia. These findings support its feasibility for AI-assisted detection of referable DR in resource-limited settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fundus camera and artificial intelligence for diabetic retinopathy screening in a São Paulo shelter.Revista da Associacao Medica Brasileira (1992) · 2026
- Performance Assessment of an Artificial Intelligence System Used for Case Detection in a Large Community-based Diabetic Retinopathy Screening Program: A Retrospective Diagnostic Accuracy Validation Study.Acta medica Philippina · 2026
- Autonomous AI-driven point-of-care screening for diabetic retinopathy compared to reading center multi-expert clinical review: results from three prospective controlled pivotal validation studies with AEYE-DS in over 1,200 patients.Frontiers in digital health · 2026
- Diabetic retinopathy screening using artificial intelligence: a comprehensive performance comparison on datasets from diverse populations and imaging modalities.Physiological measurement · 2026
- Ocular Factors Affecting AI Diagnosis in Diabetic Retinopathy Screening for Resource-Limited Regions: Cross-Sectional Study.Journal of medical Internet research · 2026
- Real-world performance of an AI system for diabetic retinopathy screeningScientific Reports · 2026