Fundus camera and artificial intelligence for diabetic retinopathy screening in a São Paulo shelter.
Julia Gimenes da Silva Dias, Gabriel Costa de Andrade, João Victor Tenório Lossio de Macedo, Andrezza Soldera Graziano
Irmandade da Santa Casa de Misericórdia de São Paulo - São Paulo (SP), Brazil. · Irmandade da Santa Casa de Misericórdia de São Paulo, Department of Ophthalmology - São Paulo (SP), Brazil.
Key points
From the paper's conclusion- The Eyer® device demonstrated operational feasibility for ocular screening by trained non-ophthalmologists. The high prevalence of media opacities and non-diabetic retinopathy conditions compromised artificial intelligence performance, highlighting the need for broader screening strategies in elderly and socially vulnerable populations.
Abstract
OBJECTIVE: The aim of this study was to evaluate the applicability of a smartphone-based portable fundus camera for diabetic retinopathy screening in elderly residents of a specialized shelter in São Paulo, Brazil, considering image acquisition by non-ophthalmologist professionals and remote case classification. METHODS: Prospective pilot interventional study at the Centro de Acolhida Especial para Idosos Morada São João. Thirty-eight participants (76 eyes) underwent retinal imaging using the Eyer® portable fundus camera (Phelcom Technologies, Brazil) by trained non-ophthalmologist professionals. Images were remotely evaluated by a retinal specialist and compared to the EyerMaps® artificial intelligence system. RESULTS: Mean age was 71.6±7.0 years; 78.9% male. Fifteen participants (39.5%) had confirmed diabetes mellitus. The specialist identified abnormalities in 70 of 76 eyes (92.1%), most frequently glaucoma suspicion, media opacity suggestive of cataract, chorioretinal changes, and age-related macular degeneration signs; no diabetic retinopathy was identified. Against the specialist's assessment of any ocular abnormality as the reference standard, the artificial intelligence system classified 22 eyes as abnormal, yielding sensitivity 25.7% (95%CI 16.9-37.0), specificity 33.3% (95%CI 9.7-70.0), positive predictive value 81.8% (95%CI 61.5-92.7), negative predictive value 3.7% (95%CI 1.0-12.5), and overall accuracy 26.3% (95%CI 17.7-37.2). CONCLUSION: The Eyer® device demonstrated operational feasibility for ocular screening by trained non-ophthalmologists. The high prevalence of media opacities and non-diabetic retinopathy conditions compromised artificial intelligence performance, highlighting the need for broader screening strategies in elderly and socially vulnerable populations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world validation of an AI-based non-mydriatic fundus camera in the diagnosis of diabetic retinopathy in Liberia.Frontiers in digital health · 2026
- Performance Assessment of an Artificial Intelligence System Used for Case Detection in a Large Community-based Diabetic Retinopathy Screening Program: A Retrospective Diagnostic Accuracy Validation Study.Acta medica Philippina · 2026
- Diabetic retinopathy screening using artificial intelligence: a comprehensive performance comparison on datasets from diverse populations and imaging modalities.Physiological measurement · 2026
- Autonomous AI-driven point-of-care screening for diabetic retinopathy compared to reading center multi-expert clinical review: results from three prospective controlled pivotal validation studies with AEYE-DS in over 1,200 patients.Frontiers in digital health · 2026
- Ocular Factors Affecting AI Diagnosis in Diabetic Retinopathy Screening for Resource-Limited Regions: Cross-Sectional Study.Journal of medical Internet research · 2026
- Real-world performance of an AI system for diabetic retinopathy screeningScientific Reports · 2026