Frontiers in Cardiovascular Medicine

Case Report: Steerable catheter- and balloon-assisted recapture of a self-expanding pulmonary valve: a novel technical solution

Anastasios Chatziantoniou, Enrico Piccinelli, Micol Rebonato, Mara Pilati, Gianfranco Butera

Case report

Key points

From the paper's conclusion
  • A steerable catheter and a balloon were used to realign the valve and recapture it in the DrySeal sheath.

Abstract

A 24-year-old female patient, with tetralogy of Fallot corrected with a transannular patch as an infant, presented with severe pulmonary valve regurgitation and right ventricle dilatation. A transcatheter pulmonary valve replacement was decided upon with a self-expanding Harmony valve. During the implantation, the valve jumped lower than expected in the main pulmonary artery and recapture was therefore attempted. A steerable catheter and a balloon were used to realign the valve and recapture it in the DrySeal sheath.