Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation

Finn Åkerström, Emmanouil Charitakis, Fahd Asaad, Serkan Saygi, Emma Svennberg, Frieder Braunschweig, Mats Jensen-Urstad, Nikola Drca

Key points

From the paper's conclusion
  • Conclusion Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.

Abstract

Background Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation (RF) catheter to guide pulmonary vein isolation (PVI) are limited. The use of impedance drop correlates with lesion formation but is hampered by cardiac and respiratory oscillation artifacts. A filtered impedance signal, average impedance drop (AID), is now available in the EnSite X EP system. The study aim was to evaluate the efficacy and safety of a novel ablation protocol based on proportional (%)AID using the TFSE catheter, including mandatory remapping to assess PVI durability. Methods Pulmonary vein isolation with the TFSE was performed in patients with atrial fibrillation (AF) undergoing first-time PVI. The ablation protocol consisted of reaching %AID ≥9 or reaching maximum allowed RF time of 20 s and 10 s anteriorly and posteriorly. Protocol-mandated remapping was performed ≥2 months following index ablation. Results The study included 26 patients (paroxysmal/persistent AF = 17/9). Protocol adherence was 93%, and first-pass isolation was achieved in 85% of left and 93% of right pulmonary veins (PVs). Mean %AID was 9.4 ± 0.98 and proportion of lesions with %AID ≥9 was 62.2 ± 15.5%. Durable PVI was observed in 93% of all PVs at remapping (88% of patients) with no ablation-related complications. Inter-lesion distance, but not %AID, was significantly different in reconnected PV segments compared to segments with durable isolation (Firth logit <0.001). Conclusion Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers