Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation
Finn Åkerström, Emmanouil Charitakis, Fahd Asaad, Serkan Saygi, Emma Svennberg, Frieder Braunschweig, Mats Jensen-Urstad, Nikola Drca
Key points
From the paper's conclusion- Conclusion Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.
Abstract
Background Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation (RF) catheter to guide pulmonary vein isolation (PVI) are limited. The use of impedance drop correlates with lesion formation but is hampered by cardiac and respiratory oscillation artifacts. A filtered impedance signal, average impedance drop (AID), is now available in the EnSite X EP system. The study aim was to evaluate the efficacy and safety of a novel ablation protocol based on proportional (%)AID using the TFSE catheter, including mandatory remapping to assess PVI durability. Methods Pulmonary vein isolation with the TFSE was performed in patients with atrial fibrillation (AF) undergoing first-time PVI. The ablation protocol consisted of reaching %AID ≥9 or reaching maximum allowed RF time of 20 s and 10 s anteriorly and posteriorly. Protocol-mandated remapping was performed ≥2 months following index ablation. Results The study included 26 patients (paroxysmal/persistent AF = 17/9). Protocol adherence was 93%, and first-pass isolation was achieved in 85% of left and 93% of right pulmonary veins (PVs). Mean %AID was 9.4 ± 0.98 and proportion of lesions with %AID ≥9 was 62.2 ± 15.5%. Durable PVI was observed in 93% of all PVs at remapping (88% of patients) with no ablation-related complications. Inter-lesion distance, but not %AID, was significantly different in reconnected PV segments compared to segments with durable isolation (Firth logit <0.001). Conclusion Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical and experimental evaluation of averaged impedance drop during radiofrequency catheter ablation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Residual impedance drop beyond ablation index is associated with atrial fibrillation recurrence after catheter ablationEuropean Heart Journal · 2026
- Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein GapsJournal of Arrhythmia · 2026
- Variability in Tissue Interface Temperature During Pulse Field Atrial Ablation: Implications for Real-Time Contact AssessmentCirculation Arrhythmia and Electrophysiology · 2026
- Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility StudyJournal of Arrhythmia · 2026
- Continuous drag ablation versus point-by-point lesion sets for pulmonary vein isolation: A systematic review and meta-analysisEuropean Heart Journal · 2026