Hands-on time, time to result, and inter-device concordance: a prospective real-life comparison of three MALDI-TOF MS systems in the routine laboratory
Isabel Klugherz, Hanna Greimel, Simone Friedl, Ivo Steinmetz, Karl Dichtl
Key points
From the paper's conclusion- This prospective study evaluated three test systems for their performance in this setting: Bruker’s Biotyper sirius (“Biotyper”), Zybio’s EXS2600 (EXS), and bioMérieux’s VITEK MS PRIME (“VITEK”) regarding hands-on and measurement times, measurement confidence and inter-device concordance in a routine diagnostic setting.
Abstract
Introduction MALDI-TOF-MS is the standard method for rapid microbial identification in microbiology laboratories. Several commercial systems are available; however, comparative data for real-life applications in routine diagnostics remain limited. This prospective study evaluated three test systems for their performance in this setting: Bruker’s Biotyper sirius (“Biotyper”), Zybio’s EXS2600 (EXS), and bioMérieux’s VITEK MS PRIME (“VITEK”) regarding hands-on and measurement times, measurement confidence and inter-device concordance in a routine diagnostic setting. Methods A total of 2,433 bacterial and yeast isolates recovered in a routine laboratory were tested in parallel on all three devices, analogous to everyday procedures, resulting in 254 different identification results covering commonly encountered species. The sample preparation time, measuring time, measurement confidence, need for and impact of re-testing, and inter-device concordance at the species level were assessed. Results While sample preparation times were comparable, Biotyper demonstrated significantly shorter measurement times. With one measurement, VITEK, EXS, and Biotyper yielded high-confidence identifications (green category) at 86.3%, 88.0%, and 92.4%, respectively. This increased to 93.2%, 95.7%, and 97.7% after repeated testing of isolates with initially lower confidence or invalid results (yellow or red category). Repeated measurements led to an improvement of 39-67% in re-analyzed isolates. The overall concordance between the three devices was high (93.9% of isolates with results for all three devices). Conclusions All three MALDI-TOF MS devices yielded confident results in a high proportion of isolates. Depending on the requirements, for example, the need for the shortest possible turnaround time or for avoiding repeated measurements, different devices may be the right solution for the individual laboratory.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance comparison of VITEK MS PRIME with VITEK MS in clinical hospital laboratory setting.Microbiology spectrum · 2026
- Applications of Machine Learning-Assisted MALDI-TOF Mass Spectrometry in Clinical Bacterial Analysis: Research Progress and Clinical Translation.ACS omega · 2026
- Laboratory characteristics associated with microbial identification quality: a national multicenter study based on external quality assessment across China.Microbiology spectrum · 2026
- Evaluation of the Zybio EXS2600 Mass Spectrometry for Accurate and Fast Identification of Filamentous Fungi Clinical Isolates.Journal of fungi (Basel, Switzerland) · 2026
- Validation of the Mycelium Transfer Method Coupled with MALDI-TOF MS for Rapid Identification of Filamentous Fungi.Pathogens (Basel, Switzerland) · 2026
- MSI-3, an improved version of the MSI application that includes both fungal and bacterial identifications for MALDI-ToF spectra.Microbiology spectrum · 2026