Performance comparison of VITEK MS PRIME with VITEK MS in clinical hospital laboratory setting.
Nina J Gao, Paul M Luethy, J Kristie Johnson
California Department of Public Health, Center for Laboratory Sciences, Richmond, California, USA. · Department of Pathology, University of Maryland Baltimore School of Medicine, Baltimore, Maryland, USA.
Key points
From the paper's conclusion- Our study evaluates the performance of the VITEK MS PRIME against its predecessor model in a high-throughput clinical diagnostic laboratory, putting the instrument to the test across a broad spectrum of microbial organisms.
Abstract
MALDI-TOF mass spectrometry has been instrumental in rapid organism identification from clinical microbiology cultures. bioMérieux's latest MALDI-TOF mass spectrometer for in vitro diagnostic laboratory use is the VITEK MS PRIME (MS PRIME). This instrument features a new internal multi-slide holder and streamlined sample acquisition. In this study, we performed a direct comparison between the MS PRIME and the VITEK MS for identification accuracy and time to result, to evaluate performance in a clinical laboratory setting. Two hundred fifty micro-organism isolates recovered from the University of Maryland Medical Center (UMMC) routine patient cultures were evaluated. Diagnostic accuracy for the MS PRIME was determined using the VITEK MS identification as the reference method. Organisms were considered correctly identified by the MS PRIME if an identification scored >99.0% confidence on at least two spots, and matched the VITEK MS identification. Slide loading and sample spot acquisition times were measured to compare functional time from slide preparation to receipt of final identification result for diagnostic reporting. The MS PRIME demonstrated high identification accuracy with 98.0% accuracy for all microbial isolates tested, except five failures from anaerobic bacteria (245/250). Average time to first sample acquisition from slide loading was 4 min 13 s for MS PRIME and 10 min 28 s for VITEK MS. Average spot acquisition time was significantly faster for MS PRIME (14 s, n = 100) compared to VITEK MS (25 s, n = 100). The MS PRIME demonstrates significantly faster time to results without significant loss in accuracy of microbial identification.IMPORTANCENew instrumentation is often marketed as improved upgrades to existing in vitro diagnostic technologies. Our study evaluates the performance of the VITEK MS PRIME against its predecessor model in a high-throughput clinical diagnostic laboratory, putting the instrument to the test across a broad spectrum of microbial organisms. Our study not only looks at the technical and analytical performance of the instrument but also examines analyst hands-on time and workflow efficiency as points of consideration when evaluating a new instrument.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hands-on time, time to result, and inter-device concordance: a prospective real-life comparison of three MALDI-TOF MS systems in the routine laboratoryFrontiers in Cellular and Infection Microbiology · 2026
- Evaluation of the Zybio EXS2600 Mass Spectrometry for Accurate and Fast Identification of Filamentous Fungi Clinical Isolates.Journal of fungi (Basel, Switzerland) · 2026
- Optimization of culture read times using full microbiology laboratory automation.Journal of clinical microbiology · 2026
- Applications of Machine Learning-Assisted MALDI-TOF Mass Spectrometry in Clinical Bacterial Analysis: Research Progress and Clinical Translation.ACS omega · 2026
- Validation of the Mycelium Transfer Method Coupled with MALDI-TOF MS for Rapid Identification of Filamentous Fungi.Pathogens (Basel, Switzerland) · 2026
- Laboratory characteristics associated with microbial identification quality: a national multicenter study based on external quality assessment across China.Microbiology spectrum · 2026