Frontiers in bioengineering and biotechnology

Sensor-integrated PMMA bone cement enables multimodal data acquisition in a load-bearing orthopedic knee spacer implants.

Nguyen Bach Tran, Laura Capogrosso, Christoph Dillitzer, Paul Morandell, Vincent Lallinger, Carina Micheler, Julia Werner, Judith Reiser,

Heinz-Nixdorf-Chair of Biomedical Electronics, School of Computation, Information and Technology, TranslaTUM, Technical University of Munich, Munich, Bavaria, Germany. · Department of Orthopaedics and Sport Orthopaedics, TUM School of Medicine and Health, Technical University of Munich University Hospital, Munich, Bavaria, Germany.

Key points

From the paper's conclusion
  • Here, we report the development of a sensorized knee spacer that transforms medical-grade PMMA bone cement from a passive structural material into an implantable multimodal sensing platform by embedding fully functional sensing and communication systems.

Abstract

INTRODUCTION: Periprosthetic joint infection (PJI) is a severe complication of total knee arthroplasty, is a leading cause of revision surgery, and is associated with significant morbidity. Two-stage exchange using antibiotic-loaded polymethylmethacrylate (PMMA) spacers remains the clinical gold standard, yet the decision to reimplant relies largely on indirect markers and clinical judgment, as no method allows continuous in situ assessment of infection resolution. Here, we report the development of a sensorized knee spacer that transforms medical-grade PMMA bone cement from a passive structural material into an implantable multimodal sensing platform by embedding fully functional sensing and communication systems. METHODS: We developed an integration strategy addressing previously limiting factors, including exothermic polymerization, mechanical load transfer, and spatial constraints, while retaining the conventional PMMA molding principle during final spacer fabrication. A miniaturized multimodal sensor unit was integrated within PMMA cement, and the functionality of the embedded camera, spectrometer, positioning, and temperature sensors following cement integration was evaluated. The mechanical behavior of the sensor-integrated spacer was assessed under adapted dynamic biomechanical loading conditions based on ISO 14879-1:2020. In vivo functionality and longitudinal multimodal data acquisition were evaluated in a porcine pilot experiment using human knee spacer dimensions. RESULTS: Stable integration of the miniaturized multimodal sensor unit within PMMA cement was achieved while maintaining sensor functionality and wireless data transmission despite the thermal, mechanical, and geometric constraints of bone cement integration. The functionality of the embedded camera, spectrometer, positioning, and temperature sensors was maintained following cement integration. In vivo functionality and longitudinal multimodal data acquisition were demonstrated over 148 h in the porcine pilot experiment. DISCUSSION: These findings demonstrate the technical feasibility of integrating active sensing systems into load-bearing bone-cement implants under the investigated experimental conditions. The SmartSpacer provides a platform for longitudinal multimodal data acquisition during the spacer interval, establishing a technical basis for future studies investigating infection-associated biomarkers.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers