Fatigue-aware alveolar stress and strain tracking: a theoretical biomechanics framework for regional mechanical dose accumulation in ventilator-induced lung injury.
Michael S Champagne
VentGuard Analytics, LLC, Webster, TX, United States.
Key points
From the paper's conclusion- FAAST (Fatigue-Aware Alveolar Stress and Strain Tracking) and its deterministic implementation, FAAST2, describe regional mechanical exposure and fatigue.
Abstract
Ventilator-induced lung injury remains incompletely explained by global bedside variables that do not resolve regional load, deformation, time, repetition, and recovery. FAAST (Fatigue-Aware Alveolar Stress and Strain Tracking) and its deterministic implementation, FAAST2, describe regional mechanical exposure and fatigue. The framework maps a shared central-airway boundary into 20 compartments, defines functional-residual-capacity-referenced volumetric strain, closes generalized constitutive stress through regional specific elastance, and tracks Time-Accumulated Stress History (TASH), the complete-cycle load-time channel, and Strain-Accumulated Strain History (SASH), the cumulative positive loading-branch deformation-work channel. A cycle-indexed Continuum of Damage Index (CDI) combines elemental load-time and deformation-work increments with inherited fatigue. CDI is irreversible and bounded; a continuous recovery-opportunity factor accelerates decay of reversible fatigue memory rather than subtracting accumulated damage. Applying selected ventilation values from Park's ALS report with declared PEEP and normal-lung mechanics produced a chronic-stability comparator with 1-h Strain-Accumulated Strain History 88.9 kPa, dynamic strain 0.194, per-cycle deformation work 0.098 kPa, recovery opportunity 0.827, and bounded steady fatigue 0.015; CDI was not extrapolated across chronic support. Relative to this comparator, 1-h regional SASH across Cases 1-4H was 99.5, 485.1, 723.8, and 1,510.4 kPa, or 1.12, 5.46, 8.14, and 17.0 times the comparator, respectively; illustrative 24-h maximum compartment CDI was 0.280, 0.478, 0.618, and 0.851. Higher tidal volume increased burden despite fewer breaths, while a preferentially loaded compliant region in heterogeneous ARDS carried the greatest burden. A matched OSCILLATE sensitivity applied reported day-1 arm settings to the same Panwar-H-anchored 20-compartment lung for 24 h. In the compartment attaining maximum CDI at 24 h, control ventilation produced Time-Accumulated Stress History 28.080 kPa·h, SASH 15,496 kPa, and maximum compartment CDI 0.596. Across declared HFOV tracheal oscillatory tidal-volume sensitivities of 0.5-2.0 mL/kg predicted body weight, TASH was approximately 31.299 kPa·h, SASH ranged from 13,513 to 54,053 kPa, and CDI ranged from 0.575 to 0.777. Modeled SASH arm equivalence occurred at 0.573 mL/kg; this is a mechanical comparison point, not an injury threshold. These are ecological sensitivity outputs, not measured HFOV tidal volumes, patient reconstructions, time-weighted exposures, mortality predictions, or evidence that pulmonary mechanics exclude hemodynamic harm. FAAST2 requires regional construct validation and biological calibration before clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Global Protective Ventilator Settings Are Not Regional Protection: A Compartment-Resolved Framework for Personalizing Mechanical Ventilation in Acute Respiratory Distress Syndrome.Journal of clinical medicine · 2026
- Lung stress mapping: an innovative technology to visualize the hidden risk of ventilator-induced lung injuryAmerican Journal of Respiratory and Critical Care Medicine · 2026
- Mechanical power of ventilation: tracking the damaging componentCritical Care · 2026
- Prognostic value of disease severity and mechanical ventilation intensity in acute respiratory distress syndrome: analysis of the LUNG SAFE cohortEuropean Respiratory Journal · 2025
- Association Between Mechanical Power and Ventilator-Associated Conditions in Patients Requiring Prolonged Mechanical Ventilation: A Multi-Database Retrospective Cohort Study.Journal of clinical medicine · 2026
- A Machine Learning Pipeline for Scalable Annotation of Patient-Ventilator Dyssynchrony from Bedside Ventilator DatamedRxiv · 2026