Test parameter selection affects the perceived antibacterial efficacy of a contact-killing coating developed to prevent prosthetic joint infection.
Julia L van Agtmaal, Cynthia Calligaro, Nihal Engin Vrana, Sander M J van Kuijk, Tim J M Welting, Jacobus J C Arts
Department of Orthopaedic Surgery, Research Institute CAPHRI, Maastricht University Medical Center, Maastricht, Netherlands. · SPARTHA Medical, Strasbourg, France.
Key points
From the paper's conclusion- In MHB, a >5 log reduction in CFU/sample was observed for the coated samples compared to the uncoated ones. In TSB, reductions ranged from 0.4 to 3.5 log. A volume of 0.5 mL and an agitation speed of 120 RPM resulted in higher reductions in CFU/sample compared to 1 mL and 60 RPM, respectively. The main source of variation was the type of broth, followed by agitation speed and inoculum volume.
Abstract
INTRODUCTION: Prosthetic joint infections occur in 1% -4% of total hip and knee arthroplasties. Many new treatment approaches to prevent bacterial adhesion and biofilm formation are under development. This study highlights the importance of methodological parameter selection and data processing in in vitro antibacterial testing by evaluating a contact-killing coating composed of poly-ε-lysine and hyaluronic acid (PEL-10/HA-1). METHODS: Two broth compositions, inoculum volumes, and agitation speeds were evaluated to test bacterial adhesion on PEL-10/HA-1-coated and uncoated titanium alloy Ti6Al4V (Ti) samples. The samples were inoculated with 1 × 107 colony-forming units (CFU)/mL of Staphylococcus aureus (ATCC 25923). The inoculum was prepared in tryptic soy broth (TSB) or Mueller-Hinton broth (MHB) in volumes of 0.5 or 1 mL. The inoculated samples were incubated for 24 h at 60 or 120 RPM (0.05 or 0.2 g, respectively). After incubation, the attached CFU per sample were counted. RESULTS: In MHB, a >5 log reduction in CFU/sample was observed for the coated samples compared to the uncoated ones. In TSB, reductions ranged from 0.4 to 3.5 log. A volume of 0.5 mL and an agitation speed of 120 RPM resulted in higher reductions in CFU/sample compared to 1 mL and 60 RPM, respectively. The main source of variation was the type of broth, followed by agitation speed and inoculum volume. DISCUSSION: In conclusion, methodological choices in seemingly standard antibacterial efficacy testing protocols significantly influence the outcomes. These results underscore the critical need to carefully select and thoroughly document parameters to ensure the effectivity and reliability of testing antibacterial approaches.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In vitro assessment of antibacterial and biocompatibility properties of a poly-ε-lysine and hyaluronic acid contact-killing coating to prevent prosthetic joint infectionPLoS ONE · 2026
- Antimicrobial efficiency of pH-responsive PAA/PAH-coated and flucloxacillin loaded titanium implants in the Galleria mellonella model.Drug delivery and translational research · 2026
- Ciprofloxacin-loaded poly(lactic acid) coating on polyurethane ureteral stents: surface structure, antibacterial performance and encrustation resistance.RSC advances · 2026
- Adjunctive intraoperative pulsed direct current with cefazolin: a rabbit proof-of-concept model for prevention of prosthetic joint infection.Frontiers in cellular and infection microbiology · 2026
- When the Medium Skews the Message: Commercial Mueller-Hinton Agar Media and Their Effect on Antibiogram Results of Imipenem, Ciprofloxacin, and Gentamicin Against Standard and Clinical Strains of Pseudomonas aeruginosa.International journal of microbiology · 2026
- Triclosan Surface Coating Inhibits Bacterial Attachment: In Vitro Assessment for Prosthetic Joint Infection Prevention.BioMed research international · 2026