From the paper's conclusion- To evaluate the effect of standalone iStent inject implantation on intraocular pressure and the medication burden in patients with pseudoexfoliative (PEX) glaucoma, and to compare these findings with results reported for open-angle glaucoma without subtype stratification.
Abstract
PURPOSE: To evaluate the effect of standalone iStent inject implantation on intraocular pressure and the medication burden in patients with pseudoexfoliative (PEX) glaucoma, and to compare these findings with results reported for open-angle glaucoma without subtype stratification. MATERIALS AND METHODS: A search was conducted in PubMed Central, the Cochrane Library, and the ClinicalTrials.gov registry using keywords; "iStent", "standalone", and "pseudoexfoliative". A total of 182 articles were identified, of which only one study met the inclusion criteria (see PRISMA Flow 2020 diagram). For comparison, the only available meta-analysis from Healey et al. [4] was used. (keywords: "iStent", "standalone", "meta-analysis") Results: Due to the limited number of eligible studies, a scoping review approach was adopted. The vast majority of screened studies did not provide data for individual subtypes of open-angle glaucoma or include outcomes from combined cataract surgery. The single included study comprised 15 eyes of 15 patients with PEX glaucoma. The followup period was 6 months. Intraocular pressure was reduced from 23.75 ±3.28 mmHg to 15.33 ±1.07 mmHg, representing a 35% reduction (p.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of iStent® Inject W in mild to moderate primary open-angle glaucoma - Indian data on pseudophakic eyes.Indian journal of ophthalmology · 2026
- Effect of iStent inject on unmedicated intraocular pressure in glaucoma: exploratory analysis of clinical trial datamedRxiv · 2026
- Efficacy and Safety of the iStent infinite Trabecular Micro-Bypass Device for Glaucoma: A Systematic Review and Meta-Analysis.Ochsner journal · 2026
- Retrospective analysis of indications and outcomes of trabecular micro-bypass stent implantation in Indian eyes.Indian journal of ophthalmology · 2026
- Preliminary Outcomes of iStent Inject vs Kahook Dual Blade Goniotomy Combined With Phacoemulsification: A Randomized Controlled Trial.Medical science monitor : international medical journal of experimental and clinical research · 2026
- One-year clinical outcomes of iStent infinite® versus iStent inject® in open-angle glaucoma.Journal francais d'ophtalmologie · 2026